FDA Adverse Event Malfunction Summary report: N

BD INSULIN SYRINGE WITH THE BD ULTRA-FINE NEEDLE

MDR report key: 9534214 · Received December 31, 2019

Report

Report Number
1920898-2019-01486
Event Type
Malfunction
Date Received
December 31, 2019
Date of Event
December 12, 2019
Report Date
January 9, 2020
Manufacturer
BD MEDICAL - DIABETES CARE
Product Code
FMF
UDI-DI
00382908468039
PMA / PMN Number
K024112
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NH, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION: CUSTOMER RETURNED (1) LOOSE 1/2CC, 8MM SYRINGE. CUSTOMER STATES THAT WHEN THE PLUNGER CAP WAS REMOVED THE PLUNGER ROD FELL OUT AND THE STOPPER WAS MISSING. THE RETURNED SYRINGE WAS EXAMINED AND EXHIBITED A MISSING STOPPER WHICH COULD CAUSE THE PLUNGER TO FALL OUT OF THE BARREL. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR BATCH# 8120882. ALL INSPECTIONS AND CHALLENGES WERE PERFORMED PER THE APPLICABLE OPERATIONS QC SPECIFICATIONS. THERE WAS ONE (1) NOTIFICATION [200756286] NOTED THAT DID NOT PERTAIN TO THE COMPLAINT. THERE WERE NO QUALITY NOTIFICATIONS OR MAINTENANCE DISPATCHES DURING THE PRODUCTION OF THIS BATCH THAT PERTAINED TO THIS DEFECT. ROOT CAUSE CANNOT BE DETERMINED.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING USE THE PLUNGER ROD FELL OUT OF BARREL WITH A BD INSULIN SYRINGE WITH THE BD ULTRA-FINE¿ NEEDLE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: IT WAS REPORTED THAT WHEN PLUNGER CAP WAS REMOVED THE PLUNGER ROD FELL OUT.

Additional Manufacturer Narrative · 1

A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING USE THE PLUNGER ROD FELL OUT OF BARREL WITH A BD INSULIN SYRINGE WITH THE BD ULTRA-FINE¿ NEEDLE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: IT WAS REPORTED THAT WHEN PLUNGER CAP WAS REMOVED THE PLUNGER ROD FELL OUT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1331639 BD INSULIN SYRINGE WITH THE BD ULTRA-FINE NEEDLE PISTON SYRINGE FMF BD MEDICAL - DIABETES CARE 328468 8120882 00382908468039

Patients

Seq Age Sex Outcome Treatment
1 Other