FDA Adverse Event Malfunction Summary report: N

HEARTMATE MOBILE POWER UNIT, NA

MDR report key: 9532279 · Received December 30, 2019

Report

Report Number
2916596-2019-06124
Event Type
Malfunction
Date Received
December 30, 2019
Date of Event
December 10, 2019
Report Date
March 19, 2020
Manufacturer
THORATEC CORPORATION
Product Code
DSQ
PMA / PMN Number
P060040
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

MANUFACTURER'S INVESTIGATION CONCLUSION: THE REPORTED EVENT OF MULTIPLE NO EXTERNAL POWER EVENTS WAS CONFIRMED VIA THE PROVIDED LOG FILE. THE LOG FILE CONTAINED DATA SPANNING 3 DAYS ((B)(6) 2019 ¿ (B)(6) 2019 PER TIME STAMP. NO EXTERNAL POWER ALARMS WERE OBSERVED ON (B)(6) 2019 AT 11:37 AND AT 21:44, BOTH WHILE THE MPU WAS IN USE. THE NO EXTERNAL POWER EVENTS APPEARED TO HAVE BEEN CAUSED BY A LOSS OF AC POWER, AS THE RSOC AND VOLTAGE VALUES STEADILY DECREASED OVER 3-4 SECONDS DURING BOTH INSTANCES. THE BACKUP BATTERY OPERATED THE SYSTEM DURING THESE TIMES AND SUPPORT TO THE PUMP WAS NOT INTERRUPTED. BOTH ALARMS RESOLVED WITHIN ONE MINUTE WHEN POWER WAS RESTORED TO THE SYSTEM. NO OTHER EVENTS REGARDING THE MPU WERE OBSERVED THROUGHOUT THE LOG FILE. THE MPU (SERIAL NUMBER UNKNOWN) WAS NOT RETURNED FOR ANALYSIS. THE ROOT CAUSE OF THE NO EXTERNAL POWER EVENTS WAS DETERMINED TO BE RELATED TO A LOSS OF AC POWER WHILE THE CONTROLLER WAS CONNECTED TO THE MPU, HOWEVER THE CAUSE FOR THE LOSS OF AC POWER WAS UNABLE TO BE CONCLUSIVELY DETERMINED THROUGH THIS ANALYSIS. NO FURTHER INFORMATION WAS PROVIDED. THE MANUFACTURER IS CLOSING THE FILE ON THIS EVENT.

Additional Manufacturer Narrative · 1

ATTEMPTS FOR THE DEVICE SERIAL NUMBER HAVE BEEN MADE, HOWEVER NO RESPONSE HAS BEEN RECEIVED. NO FURTHER INFORMATION WAS PROVIDED. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER¿S INVESTIGATION IS COMPLETED. PMA/510(K) #: P160054.

Description of Event or Problem · 1

IT WAS REPORTED THAT ROUTINE LOG FILES WERE SENT. DURING THE ANALYSIS OF THE LOG FILES IT WAS OBSERVED THAT THERE WERE MULTIPLE NO EXTERNAL POWER ALARMS WHILE ON THE MOBILE POWER UNIT(MPU). THIS WAS CAUSED BY POWER TO THE MPU BEING BRIEFLY INTERRUPTED. THIS MAY BE DUE TO THE POWER CORD COMING LOOSE FROM THE WALL OUTLET OR THE HOUSE LOSING POWER. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1325023 HEARTMATE MOBILE POWER UNIT, NA VENTRICULAR (ASSISST) BYPASS DSQ THORATEC CORPORATION 107754

Patients

Seq Age Sex Outcome Treatment
1 58 YR