HEARTMATE MOBILE POWER UNIT, NA
Report
- Report Number
- 2916596-2019-06124
- Event Type
- Malfunction
- Date Received
- December 30, 2019
- Date of Event
- December 10, 2019
- Report Date
- March 19, 2020
- Manufacturer
- THORATEC CORPORATION
- Product Code
- DSQ
- PMA / PMN Number
- P060040
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
MANUFACTURER'S INVESTIGATION CONCLUSION: THE REPORTED EVENT OF MULTIPLE NO EXTERNAL POWER EVENTS WAS CONFIRMED VIA THE PROVIDED LOG FILE. THE LOG FILE CONTAINED DATA SPANNING 3 DAYS ((B)(6) 2019 ¿ (B)(6) 2019 PER TIME STAMP. NO EXTERNAL POWER ALARMS WERE OBSERVED ON (B)(6) 2019 AT 11:37 AND AT 21:44, BOTH WHILE THE MPU WAS IN USE. THE NO EXTERNAL POWER EVENTS APPEARED TO HAVE BEEN CAUSED BY A LOSS OF AC POWER, AS THE RSOC AND VOLTAGE VALUES STEADILY DECREASED OVER 3-4 SECONDS DURING BOTH INSTANCES. THE BACKUP BATTERY OPERATED THE SYSTEM DURING THESE TIMES AND SUPPORT TO THE PUMP WAS NOT INTERRUPTED. BOTH ALARMS RESOLVED WITHIN ONE MINUTE WHEN POWER WAS RESTORED TO THE SYSTEM. NO OTHER EVENTS REGARDING THE MPU WERE OBSERVED THROUGHOUT THE LOG FILE. THE MPU (SERIAL NUMBER UNKNOWN) WAS NOT RETURNED FOR ANALYSIS. THE ROOT CAUSE OF THE NO EXTERNAL POWER EVENTS WAS DETERMINED TO BE RELATED TO A LOSS OF AC POWER WHILE THE CONTROLLER WAS CONNECTED TO THE MPU, HOWEVER THE CAUSE FOR THE LOSS OF AC POWER WAS UNABLE TO BE CONCLUSIVELY DETERMINED THROUGH THIS ANALYSIS. NO FURTHER INFORMATION WAS PROVIDED. THE MANUFACTURER IS CLOSING THE FILE ON THIS EVENT.
ATTEMPTS FOR THE DEVICE SERIAL NUMBER HAVE BEEN MADE, HOWEVER NO RESPONSE HAS BEEN RECEIVED. NO FURTHER INFORMATION WAS PROVIDED. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER¿S INVESTIGATION IS COMPLETED. PMA/510(K) #: P160054.
IT WAS REPORTED THAT ROUTINE LOG FILES WERE SENT. DURING THE ANALYSIS OF THE LOG FILES IT WAS OBSERVED THAT THERE WERE MULTIPLE NO EXTERNAL POWER ALARMS WHILE ON THE MOBILE POWER UNIT(MPU). THIS WAS CAUSED BY POWER TO THE MPU BEING BRIEFLY INTERRUPTED. THIS MAY BE DUE TO THE POWER CORD COMING LOOSE FROM THE WALL OUTLET OR THE HOUSE LOSING POWER. NO FURTHER INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1325023 | HEARTMATE MOBILE POWER UNIT, NA | VENTRICULAR (ASSISST) BYPASS | DSQ | THORATEC CORPORATION | 107754 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 58 YR |