FDA Adverse Event Injury Summary report: N

CUP: VERSAFITCUP ACETABULAR SHELL Ø 48

MDR report key: 9525226 · Received December 27, 2019

Report

Report Number
3005180920-2019-01084
Event Type
Injury
Date Received
December 27, 2019
Date of Event
November 30, 2019
Report Date
December 27, 2019
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
MEH
UDI-DI
07630030808036
PMA / PMN Number
K083116
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 1

BATCH REVIEW PERFORMED ON 4 DECEMBER 2019. LOT 165684: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 25-JAN-2017 EXPIRATION DATE: 01/10/2022. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY OTHER SIMILAR REPORTED EVENTS. ADDITIONAL IMPLANT INVOLVED: STEM: MINIMAX 01.13.102R CEMENTLESS ANATOMICAL STEM RIGHT SIZE 2 (K170845) LOT. 172015. BATCH REVIEW PERFORMED ON 4 DECEMBER 2019 LOT 172015: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 31-AUG-2017 EXPIRATION DATE: 18/08/2022. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY OTHER SIMILAR REPORTED EVENTS. CLINICAL EVALUATION PERFORMED BY MEDICAL AFFAIRS MANAGER HIP REVISION SURGERY PERFORMED 1 YEAR AND 8 MONTHS AFTER CEMENTLESS DOUBLE MOBILITY TOTAL HIP ARTHROPLASTY IN A (B)(6) YEAR OLD WOMAN. RADIOGRAPHIC IMAGE PROVIDED SHOWS SUBOPTIMAL CUP POSITION. THE REASON OF THIS CHOICE IS UNKNOWN: BONE MORPHOLOGY OR PATIENT ANATOMY MAY BE ONE REASON BUT THIS CANNOT BE CONFIRMED ON THE BASIS OF A SINGLE ANTEROPOSTERIOR X-RAY. THERE IS NO REASON TO SUSPECT A FAULTY DEVICE.

Description of Event or Problem · 1

REVISION SURGERY PERFORMED 1 YEAR AND 9 MONTHS AFTER THE PRIMARY DUE TO IMPINGEMENT (THE CUP WITH THE STEM OR THE LESSER TROCHANTER WITH THE PELVIS). THE CUP WAS ANTEVERTED (THE CAUSE IS UNKNOWN). THE SURGEON REVISED THE HEAD AND LINER. INFECTION NOT PRESENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1318799 CUP: VERSAFITCUP ACETABULAR SHELL Ø 48 ACETABULAR CUP MEH MEDACTA INTERNATIONAL SA 165684 07630030808036

Patients

Seq Age Sex Outcome Treatment
1 56 YR Required Intervention