FDA Adverse Event Injury Summary report: N

THINWALLED FEP RINGED GORE-TEX VASCULAR GRAFT

MDR report key: 951492 · Received November 19, 2007

Report

Report Number
2017233-2007-00392
Event Type
Injury
Date Received
November 19, 2007
Date of Event
January 1, 1995
Report Date
November 19, 2007
Manufacturer
W.L. GORE & ASSOCIATES, INC.
Product Code
DSY
PMA / PMN Number
K881732
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE MANUFACTURING LOT # IS UNK. THE IMPLANT DATE REMAINS UNK. SAMOTOWKA M, KAHN MB, ROBERTS AB, BAKSHI KR, FAIRMAN RM. THIN-WALLED POLYTETRAFLUOROETHYLENE GRAFT FAILURE IS A CAUSE OF AXILLARY PULLOUT SYNDROME. SURGERY 1995; 117:113-4.

Description of Event or Problem · 1

VIA SCIENTIFIC LITERATURE, THE PT PRESENTED WITH A BILATERAL EXTERNAL ILIAC ARTERY OCCLUSIONS AND A LEFT SUPERFICIAL FEMORAL ARTERY OCCLUSION. A LEFT AXILLOPROFUNDA BYPASS WAS PERFORMED. THE POSTOPERATIVE COURSE WAS UNEVENTFUL AND THE PT WAS DISCHARGED. THREE WEEKS POST PROCEDURE, THE PT DEVELOPED AN INFRACLAVICULAR PULSATILE MASS. AT OPERATION, IT WAS FOUND THAT THE GRAFT HAD RIPPED AND THAT THE SUTURE WAS INTACT ON THE ARTERY. AN INTERPOSITION GRAFT WAS PLACED. INTRAOPERATIVE CULTURES WERE NEGATIVE. THE PT HAD AN UNEVENTFUL POSTOPERATIVE COURSE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 THINWALLED FEP RINGED GORE-TEX VASCULAR GRAFT VASCULAR PROSTHESIS DSY W.L. GORE & ASSOCIATES, INC. WLG401 UNK

Patients

Seq Age Sex Outcome Treatment
1 56 YR Life Threatening