FDA Adverse Event Injury Summary report: N

MENTOR MEMORYGEL BREAST IMPLANT

MDR report key: 9513737 · Received December 23, 2019

Report

Report Number
1645337-2019-26085
Event Type
Injury
Date Received
December 23, 2019
Date of Event
August 1, 2019
Report Date
November 25, 2019
Manufacturer
MENTOR TEXAS
Product Code
FTR
UDI-DI
00081317000501
PMA / PMN Number
P030053
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

ON MAY 5, 2020, THE MENTOR FAILURE ANALYSIS LAB HAS RECEIVED THE DEVICE FOR EVALUATION. ON MAY 13, 2020, THE PRODUCT INVESTIGATION WAS COMPLETED. DEVICE INVESTIGATION SUMMARY: DURING VISUAL EVALUATION OF THE DEVICE, AN ANOMALY WAS OBSERVED ON THE ANTERIOR VIEW OF THE DEVICE CONSISTENT WITH A CREASE/FOLD. A TEAR WAS ALSO OBSERVED WITHIN THE CREASE/FOLD, MEASURING APPROXIMATELY 0.6 CM. THE EVALUATION DETERMINED THAT THE LOSS OF SHELL INTEGRITY IS CONSISTENT WITH A CREASE FOLD RUPTURE OF THE IMPLANT SHELL, WHICH IS A KNOWN INHERENT RISK OF GEL-FILLED MAMMARY PROSTHESIS. RUPTURE CAN OCCUR AT ANY TIME AFTER IMPLANTATION, BUT IT IS MORE LIKELY TO OCCUR THE LONGER THE IMPLANT IS IMPLANTED. CREATING WRINKLES OR FOLDS SHOULD BE AVOIDED DURING IMPLANTATION OR OTHER PROCEDURES AS IT CAN RESULT IN A RUPTURE IN A LATER TIME. BREAST IMPLANTS MAY ALSO SIMPLY WEAR OUT OVER TIME. MENTOR CONCLUDED THAT THE CAPSULAR CONTRACTURE IN THE PATIENT´S BREAST WAS THE RESULT OF THE BODY´S INDIVIDUAL PHYSIOLOGICAL RESPONSE TO THE IMPLANTATION OF A FOREIGN OBJECT IN SOFT TISSUE. CAPSULAR CONTRACTURE IS A KNOWN COMPLICATION ASSOCIATED WITH THESE DEVICES AND IS REFERENCED IN OUR CURRENT PRODUCT INSERT DATA SHEET. THE AMERICAN SOCIETY OF PLASTIC SURGEONS RECOMMENDS AND ENCOURAGES MEMBER SURGEONS TO ALWAYS SUBMIT BREAST IMPLANTS, CAPSULE, AND EFFUSION TO PATHOLOGY FOR EXAMINATION. AS PART OF OUR QUALITY PROCESS, THE MANUFACTURING RECORDS OF LOT 5971283 WERE REVIEWED AND THE MANUFACTURING STANDARDS WERE MET PRIOR TO THE RELEASE OF THIS LOT. EACH DEVICE IS VISUALLY INSPECTED DURING MANUFACTURING TO ENSURE THE DEVICE MEETS THE REQUIRED SPECIFICATIONS PRIOR TO SHIPMENT. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Additional Manufacturer Narrative · 0

ON (B)(6) 2020, THE PRODUCT INVESTIGATION WAS COMPLETED. A MANUFACTURING RECORD EVALUATION (MRE) WAS PERFORMED FOR THE COMPLAINT DEVICE. THE MANUFACTURING RECORD EVALUATION (MRE) WAS REVIEWED, AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE MRE VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Additional Manufacturer Narrative · 1

SINCE THE DEVICE HAS NOT BEEN RETURNED FOR ANALYSIS, NO PRODUCT FAILURE ANALYSIS CAN BE CONDUCTED, AND NO DETERMINATION OF POSSIBLE CONTRIBUTING FACTORS CAN BE MADE. AS SUCH, THE INVESTIGATION WILL BE CLOSED. IF THE COMPLAINT DEVICE IS RECEIVED IN THE FUTURE, THE INVESTIGATION WILL BE REOPENED AND CONDUCTED AS APPROPRIATE. A MANUFACTURING RECORD EVALUATION IS IN PROGRESS. ONCE COMPLETED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. REASON FOR DEVICE EXPLANT AND/OR REOPERATION: CAPSULAR CONTRACTURE, RUPTURE. CONCOMITANT PRODUCTS: 400CC MENTOR MEMORYGEL BREAST IMPLANT CATALOG: 3504004BC LOT: 5971283 SN: (B)(4). MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT A (B)(6) FEMALE PATIENT WHO UNDERWENT PRIMARY BREAST AUGMENTATION WITH TWO 400CC MENTOR MEMORYGEL BREAST IMPLANTS, SUFFERED LEFT BREAST IMPLANT RUPTURE AND LEFT BREAST CAPSULAR CONTRACTURE; BAKER GRADE UNKNOWN, POST-OPERATIVELY. AS A RESULT, THE PATIENT WAS SCHEDULED FOR IMPLANT EXPLANTATION ON (B)(6) 2020.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1308281 MENTOR MEMORYGEL BREAST IMPLANT PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED FTR MENTOR TEXAS 3504004BC 5971283 00081317000501

Patients

Seq Age Sex Outcome Treatment
1 46 YR Required Intervention