UNIFINE PENTIPS 31G X 8MM
Report
- Report Number
- 8021764-2019-00016
- Event Type
- Injury
- Date Received
- December 20, 2019
- Report Date
- August 25, 2020
- Manufacturer
- OWEN MUMFORD LIMITED
- Product Code
- FMI
- UDI-DI
- 00384703530010
- PMA / PMN Number
- K152339
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
THIRD PARTY MANUFACTURER INVESTIGATED SAMPLE PROVIDED BY PATIENT UNDER MAGNIFICATION. THE DEVICE SHOWED EVIDENCE OF REUSE, WITH FOLDING OF THE GLUE AND CANNULA, WHICH INDICATED THE NEEDLE BROKE DUE TO MISUSE. PATIENT ADMITTED REUSING NEEDLE MULTIPLE TIMES BEFORE THE BREAK. REUSE OF A NEEDLE WILL CAUSE IT TO BREAK. INVESTIGATION REPORTS ATTACHED. ATTACHMENT: [GBC-19-05894 - E53-4460.PDF, SR_1_1405405338_OM_UPDATE.PDF].
PATIENT HAS RETURNED PENTIP THAT SHE USED BUT HAS NOT YET RETURNED THE MEDICAL QUESTIONNAIRE WITH DETAILS. THE PATIENT ADMITS REUSING HER PEN NEEDLES 2-3 TIMES PER INJECTION BEFORE CHANGING PEN NEEDLES. USED PENTIP HUB (MINUS THE NEEDLE) HAS BEEN RETURNED TO OUR MANUFACTURER FOR EVALUATION.
CUSTOMER CALLED IN TO LET US KNOW THAT SHE HAD GIVEN HERSELF AN INJECTION AND WHEN SHE WENT TO REMOVE THE NEEDLE FROM HER SKIN, THE NEEDLE REMAINED IN HER SKIN. SHE WENT TO THE EMERGENCY ROOM WHERE THEY DID A SONOGRAM AND LOCATED THE NEEDLE, HOWEVER, WOULD NOT ATTEMPT TO REMOVE THE NEEDLE. THE ER ADVISED HER TO WAIT AND SEE IF IT EXPELLED ITSELF BECAUSE GOING TO REMOVE IT SURGICALLY WOULD CAUSE MORE DAMAGE TO HER. THE PATIENT ADMITTED TO USING HER PEN NEEDLES 2-3 TIMES BEFORE CHANGING THE NEEDLE, IN SPITE OF INSTRUCTIONS TO USE THE NEEDLE ONLY ONE TIME. SHE WAS INSTRUCTED AGAIN TO FOLLOW THE INSTRUCTIONS, TO USE THE PEN NEEDLES ONE TIME AND DISCARD IN A SHARPS CONTAINER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1295872 | UNIFINE PENTIPS 31G X 8MM | PEN NEEDLES | FMI | OWEN MUMFORD LIMITED | AN 3530 | 1703722.U038 | 00384703530010 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 82 YR | Other |