FDA Adverse Event Injury Summary report: N

TOPCO

MDR report key: 9506946 · Received December 20, 2019

Report

Report Number
1422443-2019-00007
Event Type
Injury
Date Received
December 20, 2019
Report Date
December 20, 2019
Manufacturer
HANGZHOU SEJOY ELECTRONICS & INSTRUMENTS CO, LTD
Product Code
FLL
Adverse Event
Yes
Report Source
Distributor report
Reporter Location
IN, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DIGITAL THERMOMETER WAS READING LOW. THE THERMOMETER WAS USED ON THE REPORTER'S (B)(6)-YEAR-OLD SON AND DISPLAYED A READING OF 94.7 REPEATEDLY.. THE MOTHER WAS CONCERNED AND EXPRESSED THE CHILD HAD SYMPTOMS, SUCH AS BEING SWEATY, AND THE CHILD WAS TAKEN TO THE ER. THE ER MEASURED THE CHILD'S TEMPERATURE TO BE 104 F RECTALLY. THE REPORTER TRIED THE UNIT IN HER ARMPIT, AND GOT A READING OF 96.8 REPEATEDLY. THE UNIT WAS IN SERVICE FOR LESS THAN A YEAR. THE DEVICE WAS RETURNED AND EVALUATED AND WAS FOUND TO BE FUNCTIONING PROPERLY AND APPEARED TO BE GIVING ACCURATE TEMPERATURE READINGS. THE DEVICE WILL BE SENT BACK TO THE MANUFACTURER FOR THEIR VERIFICATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1295167 TOPCO 10 SECOND FLEXIBLE TIP DIGITAL THERMOMETER FLL HANGZHOU SEJOY ELECTRONICS & INSTRUMENTS CO, LTD 87-736-001 59776010186163

Patients

Seq Age Sex Outcome Treatment
1 3 YR Hospitalization