FDA Adverse Event Malfunction Summary report: N

TYPE-S HEAD EXTENSION

MDR report key: 950580 · Received April 20, 2007

Report

Report Number
1932738-2007-00001
Event Type
Malfunction
Date Received
April 20, 2007
Date of Event
March 26, 2007
Report Date
April 11, 2007
Manufacturer
CIVCO MEDICAL SOLUTIONS, MED-TEC INC.
Product Code
IWY
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CLINICIAN/HOSPITAL GLUED JOINT TOGETHER PRIOR TO RETURNING FOR EVALUATION WHICH MADE EVALUATION OF THE ORIGINAL JOINT ADHESIVE IMPOSSIBLE. PATIENT REPOSITIONED SELF PUTTING EXCESSIVE PRESSURE ON WORN DEVICE. INSTRUCTIONS FOR USE DESCRIBES AND DIAGRAMS CORRECT REPOSITIONING PROCESS OF PATIENT/CLINICIAN. INSTRUCTIONS FOR USE INDICATE TO INSPECT FOR DAMAGE/WEAR PRIOR TO USE AND NOT TO USE A DAMAGED DEVICE.

Description of Event or Problem · 1

CIVCO WAS NOTIFIED THAT A TYPE-S HEAD EXTENSION JOINT BROKE NEAR THE LOCKING MECHANISM AND ALLOWED THE PATIENT TO FALL. THE PATIENT WAS LYING FLAT ON THEIR BACK AS THEY ADJUSTED THEMSLEF OF THE TYPE-S HEAD EXTENSION. THE TYPE-S HEAD EXTENSION BROKE AND THE PATIENT FELL ABOUT 2-3 FEET. THE PATIENT WAS EVALUATED BY X-RAY AND NO INJURY WAS IDENTIFIED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TYPE-S HEAD EXTENSION CARBON FIBER COUCH INSERT IWY CIVCO MEDICAL SOLUTIONS, MED-TEC INC. 20-CFHN-SUB2-V 2901

Patients

Seq Age Sex Outcome Treatment
1 YR Other