FDA Adverse Event Malfunction Summary report: N

BIOFIRE FILMARRAY V.1.7 RESPIRATORY PANEL

MDR report key: 9503258 · Received December 20, 2019

Report

Report Number
9503258
Event Type
Malfunction
Date Received
December 20, 2019
Date of Event
August 15, 2019
Report Date
December 17, 2019
Manufacturer
BIOFIRE DIAGNOSTICS, LLC
Product Code
QDS
Product Problem
Yes
Report Source
User Facility report
Reporter Location
MD
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Description of Event or Problem · 1

RSV. WE OBSERVED AN UNEXPECTED UPTICK IN RSV POSITIVES BEGINNING (B)(6) THIS YEAR USING THE FILMARRAY V.1.7 RESPIRATORY PANEL, POUCH LOT NUMBER 671519 (FROM BOX (B)(4)). WE TEMPORARILY STOPPED USING THAT LOT, THE UNEXPECTED RSV POSITIVES DID NOT CONTINUE, BUT NO FURTHER INVESTIGATION TOOK PLACE AT THAT TIME. IN (B)(6) THIS YEAR THE LOT IN QUESTION, 671519, WAS RETURNED TO CLINICAL USE. IN (B)(6) WE OBSERVED A SECOND UNEXPECTED SERIES OF RSV POSITIVES IN SIX SPECIMENS FROM PATIENTS (B)(6). AFTER A LOOKBACK, THREE ADDITIONAL RSV POSITIVES WERE FOUND IN PATIENTS (B)(6) FROM LATE (B)(6) PATIENT (B)(6) SPECIMENS WERE RETESTED USING POUCH LOT 940119 AND WERE RSV NEGATIVE. TO RESOLVE THE DISCREPANCY, THE SIX SPECIMENS FROM PATIENTS (B)(6) WERE TESTED A THIRD TIME WITH LUMINEX ARIES FLU A/B & RSV ASSAY AND WERE NEGATIVE. THE THREE SPECIMENS FROM PATIENTS (B)(6) WERE NOT AVAILABLE FOR ADDITIONAL TESTING. THE LIMIT OF DETECTION (LOD) WAS COMPARED BETWEEN THE FILMARRAY LOT 671519, LOT 940119, AND LUMINEX. STOCK OF CELL CULTURE GROWN RSV OBTAINED FROM AN (B)(6) RESEARCH LABORATORY AND USED FOR SERIAL DILUTIONS ABOVE AND BELOW LIMIT OF DETECTION (LOD) STATED BY BIOFIRE (2 TCID50/ML). WE TESTED 10-FOLD SERIAL DILUTIONS BETWEEN 0.01X LOD AND 1000X LOD. THE LAB QUANTIFIED THE RSV QUANTITY BY A TAQMAN ASSAY USING DNA PLASMIDS AND GENERATING A STANDARD CURVE. AN ABERRANT RSV POSITIVE WAS DETECTED AT 0.01 X LOD FOR LOT 671519 CONFIRMING TO US THAT THE LOT IS DEFECTIVE AND RESULTING IN FALSE POSITIVE RSV. INSTRUMENT: BIOFIRE DIAGNOSTICS, LLC (B)(4). (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1295614 BIOFIRE FILMARRAY V.1.7 RESPIRATORY PANEL FILMARRAY V.1.7 RESPIRATORY PANEL QDS BIOFIRE DIAGNOSTICS, LLC V.1.7 RESPIRATORY PANEL

Patients

Seq Age Sex Outcome Treatment
1