FDA Adverse Event Malfunction Summary report: N

EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE

MDR report key: 9502611 · Received December 20, 2019

Report

Report Number
8010047-2019-04500
Event Type
Malfunction
Date Received
December 20, 2019
Date of Event
November 7, 2019
Report Date
January 30, 2020
Manufacturer
OLYMPUS MEDICAL SYSTEMS CORP.
Product Code
FDS
UDI-DI
04953170305290
PMA / PMN Number
K131780
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE SUBJECT DEVICE HAS NOT BEEN RETURNED TO OMSC BUT WAS RETURNED TO (B)(4). (B)(4) SENT THE DEVICE TO A THIRD PARTY LABORATORY FOR MICROBIOLOGICAL TESTING. AS A RESULT OF THE TESTING, THE SAMPLE COLLECTED FROM THE ALL CHANNELS OF THE DEVICE TESTED POSITIVE FOR COAGULASE-NEGATIVE STAPHYLOCOCCI (1 CFU/ENDOSCOPE). THE TESTING RESULT CLEARED THE FRENCH GUIDELINE. THE EXACT CAUSE OF THE REPORTED EVENT COULD NOT BE CONCLUSIVELY DETERMINED AT THIS TIME. IF ADDITIONAL INFORMATION BECOMES AVAILABLE, THIS REPORT WILL BE SUPPLEMENTED.

Additional Manufacturer Narrative · 1

THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO PROVIDE ADDITIONAL INFORMATION. OMSC REVIEWED THE MANUFACTURING HISTORY (DHR) OF THE SUBJECT DEVICE AND CONFIRMED NO IRREGULARITY. THE EXACT CAUSE OF THE REPORTED EVENT COULD NOT BE CONCLUSIVELY DETERMINED.

Description of Event or Problem · 1

OLYMPUS MEDICAL SYSTEMS CORP. (OMSC) WAS INFORMED THAT AS A RESULT OF ROUTINE MICROBIOLOGICAL TESTING BY THE USER FACILITY, FOLLOWING MICROBES WERE DETECTED FROM THE SAMPLE COLLECTED FROM THE SUBJECT DEVICE. ENTEROBACTERIE KES (COUNTLESS/150ML). THE DEVICE HAD BEEN REPROCESSED WITH A NON-OLYMPUS AUTOMATED ENDOSCOPE REPROCESSOR, SOLUSCOPE SERIE4, USING PERACETIC ACID. THERE WAS NO REPORT OF INFECTION ASSOCIATED WITH THIS REPORT.