ARCHITECT C4000 ANALYZER
Report
- Report Number
- 1628664-2019-00770
- Event Type
- Malfunction
- Date Received
- December 20, 2019
- Report Date
- February 11, 2020
- Manufacturer
- ABBOTT MANUFACTURING INC
- Product Code
- JJE
- PMA / PMN Number
- K981791
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
ON JAN 17, 2020 THE SUSPECT MEDICAL DEVICE WAS UPDATED FROM LN 03P68-31 LOT 69365UN18 TO ARCHITECT C4000 ANALYZER LN 02P24-40 SN (B)(6) , BOTH MANUFACTURERED IN IRVING, TEXAS. SECTIONS D1, D2, D4 WERE UPDATED TO DOCUMENT THIS CHANGE. FURTHER INVESTIGATION OF THE CUSTOMER ISSUE INCLUDED A REVIEW OF THE COMPLAINT TEXT, A SEARCH FOR SIMILAR COMPLAINTS, REVIEW OF INSTRUMENT LOGS, A REVIEW OF LABELING AND INSTRUMENT SERVICE. NO ADVERSE TREND WAS IDENTIFIED FOR THE CUSTOMER'S ISSUE. LABELING WAS REVIEWED AND FOUND TO BE ADEQUATE. THE FIELD SERVICE REPRESENTATIVE INSPECTED THE INSTRUMENT, DRAINED THE DEGASSER CANNISTER, CLEANED THE WATER BATH, MANUALLY CLEANED THE CUVETTES, REBUILT THE VACUUM PUMP, AND REPLACED A DIRTY/CLOGGED WATER BATH CHECK VALVE. THE ISSUE WAS RESOLVED BY REPLACING THE CHECK VALVE, WATER BATH FILL (PART NUMBER 7-203638-01). BASED ON ALL AVAILABLE INFORMATION AND ABBOTT DIAGNOSTICS' COMPLAINT INVESTIGATION NO PRODUCT DEFICIENCY WAS IDENTIFIED.
ALL AVAILABLE PATIENT INFORMATION WAS INCLUDED. ADDITIONAL PATIENT DETAILS ARE NOT AVAILABLE. AN EVALUATION IS IN PROCESS. A FINAL REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE.
THE CUSTOMER OBSERVED FALSELY ELEVATED MAGNESIUM RESULT ON THE ARCHITECT C4000 ANALYZER. THE FOLLOWING DATA WAS PROVIDED: INITIAL 9.5 MG/DL, REPEAT 2.1. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1300728 | ARCHITECT C4000 ANALYZER | AUTOMATED CHEMISTRY ANALYZER | JJE | ABBOTT MANUFACTURING INC | 69365UN18 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | ARCHITECT C4000 ANALYZER.| LN 02P24-40 SN (B)(6) .| LN 03P68-31 LOT 69365UN18.| MAGNESIUM REAGENTS. |