BD SYRINGE 0.3ML 31GA 6MM HALFUNIT
Report
- Report Number
- 1920898-2019-01448
- Event Type
- Malfunction
- Date Received
- December 19, 2019
- Date of Event
- December 4, 2019
- Report Date
- January 8, 2020
- Manufacturer
- BD MEDICAL - DIABETES CARE
- Product Code
- FMF
- UDI-DI
- 00382903249107
- PMA / PMN Number
- K024112
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
H.6. INVESTIGATION: CUSTOMER RETURNED (5) LOOSE 3/10CC, 6MM SYRINGES. CUSTOMER STATES THAT MARKINGS ARE OFF ON THE BARREL, PRINTINGS ARE NOT LINED UP. ALL RETURNED SYRINGES WERE TESTED USING THE PLUG GAUGE AND ALL SCALE MARKING PLACEMENTS FALL WITHIN SPECIFICATIONS. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR BATCH# 9070627. ALL INSPECTIONS AND CHALLENGES WERE PERFORMED PER THE APPLICABLE OPERATIONS QC SPECIFICATIONS. THERE WERE FIVE (5) NOTIFICATIONS [200819499, 200818801, 200819730, 200819826, 200818692] NOTED FOR SCRATCH PRINT. THERE WAS ONE (1) NOTIFICATION [200818447] NOTED FOR INK RINGS. ROOT CAUSE CANNOT BE DETERMINED AT THIS TIME AS THE ISSUE IS UNCONFIRMED. H3 OTHER TEXT : SEE H.10.
IT HAS BEEN REPORTED THAT THE BD¿ SYRINGE 0.3ML 31GA 6MM HALFUNIT HAS BEEN FOUND EXPERIENCING SCALE MARKING ISSUES DURING USE. THE FOLLOWING HAS BEEN PROVIDED BY THE INITIAL REPORTER: IT WAS REPORTED THAT THE CONSUMER STATED THAT MARKINGS ARE OFF ON THE BARREL, PRINTINGS ARE NOT LINED UP. VERBATIM: ISSUE: PARENT OF A CONSUMER REPORTED MARKINGS ARE OFF ON THE BARREL, PRINTINGS ARE NOT LINED UP. SHE STATED SHE HAS TO GIVE HER SON 5.5 UNITS OF INSULIN IN HER SONS TRESIBA, SHE CAN ESSENTIALLY DRAW 6.0 UNIT. SAMPLE AVAILABLE INCIDENT DATE-UNKNOWN; OCCURRENCE- UNKNOWN; ITEM# 324910; LOT# 9070627; EXPIRATION DATE-03-31-2024.
DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT HAS BEEN REPORTED THAT THE BD¿ SYRINGE 0.3ML 31GA 6MM HALFUNIT HAS BEEN FOUND EXPERIENCING SCALE MARKING ISSUES DURING USE. THE FOLLOWING HAS BEEN PROVIDED BY THE INITIAL REPORTER: IT WAS REPORTED THAT THE CONSUMER STATED THAT MARKINGS ARE OFF ON THE BARREL, PRINTINGS ARE NOT LINED UP. VERBATIM: ISSUE: PARENT OF A CONSUMER REPORTED MARKINGS ARE OFF ON THE BARREL, PRINTINGS ARE NOT LINED UP. SHE STATED SHE HAS TO GIVE HER SON 5.5 UNITS OF INSULIN IN HER SONS TRESIBA, SHE CAN ESSENTIALLY DRAW 6.0 UNIT. SAMPLE AVAILABLE INCIDENT DATE-UNKNOWN, OCCURRENCE- UNKNOWN, ITEM# 324910, LOT# 9070627; EXPIRATION DATE-03-31-2024.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1294663 | BD SYRINGE 0.3ML 31GA 6MM HALFUNIT | PISTON SYRINGE | FMF | BD MEDICAL - DIABETES CARE | 324910 | 9070627 | 00382903249107 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |