FDA Adverse Event Malfunction Summary report: N

BD VENFLON¿ PRO SAFETY SHIELDED IV CATHETER

MDR report key: 9496403 · Received December 19, 2019

Report

Report Number
8041187-2019-01024
Event Type
Malfunction
Date Received
December 19, 2019
Date of Event
December 4, 2019
Report Date
December 30, 2019
Manufacturer
BECTON DICKINSON MEDICAL (SINGAPORE)
Product Code
FOZ
PMA / PMN Number
N/A
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AU
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY: ONE PHOTO AND ONE VIDEO WAS RECEIVED BY OUR QUALITY TEAM FOR EVALUATION. UPON VISUAL INSPECTION OF THE VIDEO IT WAS OBSERVED THAT THE VALVE UNDER THE PORT HAD MOVED WHICH COULD BE THE REASON FOR THE LEAKAGE; THEREFORE, THE INCIDENT COULD BE VERIFIED. A DEVICE HISTORY RECORD REVIEW FOUND NO NON-CONFORMANCE'S ASSOCIATED WITH THIS ISSUE DURING PRODUCTION OF THIS BATCH. BASED ON THE QUALITY TEAM'S INVESTIGATION, THE ROOT CAUSE OF THE LEAKAGE IS DUE TO THE INJECTION VALVE MOVING WITHIN THE CANNULA HUB. A TREND FOR THE MOVING INJECTION VALVE HAS BEEN OBSERVED FOR THIS PRODUCT LINE. A PROJECT HAS BEEN STARTED TO FURTHER DETERMINE THE REASON FOR THE MOVING INJECTION VALVE IN THE VENFLON PRO SAFETY SO THAT A FIX CAN BE MADE TO FURTHER PREVENT THIS ISSUE. CAPA#(B)(4) HAS BEEN INITIATED TO ADDRESS THIS ISSUE.

Description of Event or Problem · 0

IT WAS REPORTED THAT LEAKAGE OCCURRED DURING USE WITH A BD VENFLON¿ PRO SAFETY SHIELDED IV CATHETER. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM GERMAN TO ENGLISH: RETURN FLOW VIA THE INJECTION VALVE. BLOOD AND INFUSION RETURNS TO THE INJECTION VALVE - VALVE MECHANISM IS DEFECTIVE.

Additional Manufacturer Narrative · 1

(B)(6). A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT LEAKAGE OCCURRED DURING USE WITH A BD VENFLON¿ PRO SAFETY SHIELDED IV CATHETER. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM (B)(6) TO ENGLISH: RETURN FLOW VIA THE INJECTION VALVE. BLOOD AND INFUSION RETURNS TO THE INJECTION VALVE - VALVE MECHANISM IS DEFECTIVE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1294542 BD VENFLON¿ PRO SAFETY SHIELDED IV CATHETER INTRAVASCULAR CATHETER FOZ BECTON DICKINSON MEDICAL (SINGAPORE) 9184022

Patients

Seq Age Sex Outcome Treatment
1 Other