FDA Adverse Event Injury Summary report: N

HEARTMATE GOGEAR SHOWER BAG

MDR report key: 9494875 · Received December 19, 2019

Report

Report Number
2916596-2019-05722
Event Type
Injury
Date Received
December 19, 2019
Date of Event
December 2, 2019
Report Date
March 25, 2020
Manufacturer
THORATEC CORPORATION
Product Code
DSQ
UDI-DI
00813024011149
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

MANUFACTURER'S INVESTIGATION CONCLUSION: THE REPORT OF DAMAGE COULD NOT BE CONFIRMED. THE ACCOUNT REPORTED THAT THE PATIENT¿S SHOWER BAG HAD A HOLE. THE PATIENT WAS SHOWERING AND THEIR BATTERIES GOT WET. THE PATIENT¿S SHOWER BAG WAS REPLACED AT THEIR APPOINTMENT ON (B)(6) 2019. THE HOSPITAL REQUESTED A REPLACEMENT. MULTIPLE REQUESTS FOR ADDITIONAL INFORMATION REGARDING THE PRODUCT¿S RETURN WERE SENT TO THE CUSTOMER; HOWEVER, NO ADDITIONAL INFORMATION WAS RECEIVED. NO PRODUCT WAS AVAILABLE FOR INVESTIGATION. THE HMII IFU AND PATIENT HANDBOOK STATE THAT ALL WEAR AND CARRY ACCESSORIES SHOULD BE INSPECTED PERIODICALLY DURING USE. IF AN ACCESSORY APPEARS DAMAGED OR WORN, DO NOT USE IT AND CALL YOUR HOSPITAL CONTACT FOR A REPLACEMENT. FURTHERMORE, THE HM SHOWER BAG IFU STATES THAT THE SYSTEM CONTROLLER OR BATTERIES SHOULD NEVER BE EXPOSED TO WATER. NO FURTHER INFORMATION WAS PROVIDED. THE MANUFACTURER IS CLOSING THE FILE ON THIS EVENT.

Additional Manufacturer Narrative · 1

INFORMATION WAS NOT PROVIDED. PMA/510(K) #: P060040; P160054. NO FURTHER INFORMATION WAS PROVIDED. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER¿S INVESTIGATION IS COMPLETED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THERE WAS A HOLE IN THE SHOWER BAG. PATIENT WAS SHOWERING AND HIS BATTERIES GOT WET. THE SHOWER BAG WAS REPLACED AT THE APPOINTMENT (B)(6) 2019; THE HOSPITAL REQUESTED TO REPLACE THEIR SHELF STOCK.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1288454 HEARTMATE GOGEAR SHOWER BAG VENTRICULAR (ASSISST) BYPASS DSQ THORATEC CORPORATION 104232 6940970 00813024011149

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention