REVERSE SHOULDER SYSTEM GLENOSPHERE 36XØ24.5
Report
- Report Number
- 3005180920-2019-01075
- Event Type
- Injury
- Date Received
- December 19, 2019
- Date of Event
- November 28, 2019
- Report Date
- December 19, 2019
- Manufacturer
- MEDACTA INTERNATIONAL SA
- Product Code
- PHX
- UDI-DI
- 07630040706568
- PMA / PMN Number
- K170452
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SZ
- Reporter Occupation
- 003
Narratives
BATCH REVIEW PERFORMED ON 6 DECEMBER 2019. LOT 189744: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 12-MAR-2019. EXPIRATION DATE: 2024-03-02. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY OTHER SIMILAR REPORTED EVENTS. ADDITIONAL IMPLANT INVOLVED: REVERSE SHOULDER SYSTEM 04.01.0151 GLENOID BASEPLATE Ø24.5X15 (K170452) LOT. 1904752. BATCH REVIEW PERFORMED ON 6 DECEMBER 2019. LOT 1904752: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 12-SEP-2019. EXPIRATION DATE: 2024-09-01. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY OTHER SIMILAR REPORTED EVENTS. CLINICAL EVALUATION PERFORMED BY MEDICAL AFFAIRS DIRECTOR. FEW DAYS AFTER PRIMARY RSA THE GLENOID COMPONENT DISLOCATED AND THE SCAPULAR BONE WAS FRACTURED. THIS IS MOST PROBABLY DUE TO THE WEAK QUALITY OF THE BONE. THE IMPLANTS SEEM TO HAVE BEEN CORRECTLY POSITIONED AND THERE IS NO DISCERNIBLE HINT OF A DEVICE MALFUNCTION.
ADDITIONAL SURGERY PERFORMED 7 DAYS AFTER THE PRIMARY DUE TO DISLOCATION OF THE BASEPLATE AND BREAKAGE OF THE POSTERIOR HALVE PORTION OF THE GLENOID. THE PATIENT HAS A WEAK BONE. NO INFORMATION IF IT WAS CAUSED BY TRAUMA. THE GLENOID BASEPLATE, LINER, SCREWS AND METAPHYSIS WERE EXPLANTED WITHOUT IMPLANTING NEW IMPLANTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1287593 | REVERSE SHOULDER SYSTEM GLENOSPHERE 36XØ24.5 | REVERSE GLENOSPHERE | PHX | MEDACTA INTERNATIONAL SA | 189744 | 07630040706568 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |