FDA Adverse Event Injury Summary report: N

THINWALLED FEP RINGED GORE-TEX VASCULAR GRAFT

MDR report key: 949131 · Received November 16, 2007

Report

Report Number
2017233-2007-00389
Event Type
Injury
Date Received
November 16, 2007
Date of Event
January 1, 1995
Report Date
November 16, 2007
Manufacturer
W.L. GORE & ASSOCIATES, INC.
Product Code
DSY
PMA / PMN Number
K881732
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE MANUFACTURING LOT # IS UNKNOWN. THE IMPLANT DATE REMAINS UNKNOWN. SEMOTOWKA M, KAHN MB, ROBERTS AB, BAKSHI KR, FAIRMAN RM, THIN-WALLED POLYTETRAFLUOROETHYLENE GRAFT FAILURE IS A CAUSE OF AXILLARY PULLOUT SYNDROME. SURGERY 1995; 117:113-4.

Description of Event or Problem · 1

VIA SCIENTIFIC LITERATURE, AN AXILLOBIFEMORAL BYPASS PROCEDURE WAS PERFORMED FOR THE TREATMENT OF SEVERE AORTOILIAC DISEASE. THE OPERATION AND POSTOPERATIVE COURSE WERE UNEVENTFUL. THREE WEEKS POST IMPLANT, THE PATIENT DEVELOPED COMPLICATIONS WHICH INCLUDED LEFT INFRACLAVICULAR PAIN AND SWELLING. THE PATIENT WAS TAKEN INTO SURGERY AND THE INFRACLAVICULAR REGION WAS EXPLORED. IT WAS NOTED THAT THE RING OF THE GRAFT REMAINED SUTURED TO THE AXILLARY ARTERY AND THE GRAFT ITSELF DISRUPTED JUST DISTAL TO THE SUTURE LINE. AN INTERPOSITION GRAFT WAS PLACED BETWEEN THE SUBCLAVIAN ARTERY AND THE DISRUPTED SEGMENT OF THE ORIGINAL GRAFT. CULTURES OBTAINED FROM THE ANASTOMOTIC SITE WERE NEGATIVE. THE PATIENT HAD AN UNEVENTFUL POSTOPERATIVE COURSE AND WAS DISCHARGED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 THINWALLED FEP RINGED GORE-TEX VASCULAR GRAFT VASCULAR PROSTHESIS DSY W.L. GORE & ASSOCIATES, INC. WLG401 UNK

Patients

Seq Age Sex Outcome Treatment
1 62 YR Life Threatening