ADVANCED PERFUSION SYSTEM 1
Report
- Report Number
- 1828100-2019-00665
- Event Type
- Malfunction
- Date Received
- December 18, 2019
- Date of Event
- November 27, 2019
- Report Date
- May 12, 2020
- Manufacturer
- TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
- Product Code
- DTQ
- PMA / PMN Number
- K172220
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AL, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THE REPORTED COMPLAINT WAS CONFIRMED. IF ADDITIONAL INFORMATION BECOMES AVAILABLE ON THIS COMPLAINT THAT WOULD ALTER THE FACTS AND/OR CONCLUSION, A SUPPLEMENTAL REPORT WILL BE FILED ACCORDINGLY.
UPDATED BLOCK: D10.
PER DATA LOG ANALYSIS, MULTIPLE MODULES/PUMPS ARE REPORTING INTERMITTENT '5 VOLTS (V) SUPPLY VOLTAGE TEST FAILURE' EVENTS WHICH WILL CAUSE A RED X TO APPEAR ON THE MODULE. THIS WILL ALSO CAUSE AN ALARM TO SOUND. THESE ERRORS HAPPENED ON 27-NOV-2019, 02-DEC-2019, AND 04-DEC-2019 WHEN THE LOGS WERE EXPORTED. BELOW IS THE LIST OF MODULES WITH 5V ERRORS. DATA TRANSFER 01395. LARGE ROLLER PUMP 03737. LARGE ROLLER PUMP 03740. LARGE ROLLER PUMP 03747. SMALL ROLLER PUMP 01661. THE 5V SUPPLY CAME FROM THE POWER MANAGER TO THE MODULES/PUMPS THROUGH THE LEFT AND RIGHT NETWORK INTERFACE CONTROLLER (NIC) BOARDS. IF ALL OF THESE MODULES/PUMPS ARE CONNECTED TO THE SAME NIC, THE ISSUE COULD BE THE NIC OR THE POWER CABLE CONNECTING TO IT. THE LOG CONFIRMS THE COMPLAINT. DURING LABORATORY ANALYSIS, THE PRODUCT SURVEILLANCE TECHNICIAN OBSERVED THE MEASURED VALUES ACROSS RESISTORS OL K¿, OL K¿ AND OL K¿ TO BE OPEN, LOSING CURRENT. SEVERAL RESISTORS ON THE POWER MANAGER BOARD WERE ALSO FOUND TO BE CHARRED.
THE FIELD SERVICE REPRESENTATIVE (FSR) VERIFIED THAT THERE WERE RED XS ON THE UNITS THAT WERE CONNECTED TO THE LEFT SIDE NETWORK INTERFACE CONTROLLER (NIC) PRINTED CIRCUIT BOARD (PCB). HE REPLACED THE POWER MANAGER PCB, LEFT CAN NIC, AND LEFT POWER NIC CABLES. THE UNIT OPERATED TO THE MANUFACTURER'S SPECIFICATIONS.
IT WAS REPORTED THAT DURING SET-UP OF THE DEVICE FOR A CARDIOPULMONARY BYPASS (CPB) PROCEDURE, THE CENTRAL CONTROL MONITOR (CCM) INTERMITTENTLY DISPLAYED MULTIPLE RED X'S. AS A RESULT, AN ALTERNATE DEVICE WAS EMPLOYED. THE SURGICAL PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WAS NO DELAY, NO BLOOD LOSS, NOR ADVERSE CONSEQUENCES TO THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1285978 | ADVANCED PERFUSION SYSTEM 1 | CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1 | DTQ | TERUMO CARDIOVASCULAR SYSTEMS CORPORATION | 801763 |