FDA Adverse Event Injury Summary report: N

AVID MEDICAL

MDR report key: 9483558 · Received December 17, 2019

Report

Report Number
1047429-2019-00019
Event Type
Injury
Date Received
December 17, 2019
Date of Event
August 22, 2019
Report Date
December 17, 2019
Manufacturer
AVID MEDICAL, INC. A MEDICAL ACTION IND. COMPANY
Product Code
LRO
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 1

COMPLAINT INVESTIGATION OPENED; HOWEVER, LIMITED INVESTIGATION DUE TO LACK OF APPLICABLE LOT NUMBER OR SAMPLE. NO SIMILAR COMPLAINTS HAVE BEEN REPORTS FOR ANY KITS. STERILIZATION RECORDS FIND NO FAILURES. ADDITIONALLY THE END-USER INDICATES OTHER KITS MANUFACTURED BY ALTERNATE SUPPLIERS WERE USED IN PROCEDURE AS WELL AS POSSIBLE CONTRIBUTORS.

Description of Event or Problem · 1

(B)(6) IS THE MANUFACTURER OF A CUSTOM PROCEDURE TRAY (REF: 140-01 BAL CATH PREP PK W/SPEC TRAP). AVID MEDICAL RECEIVED A COMPLAINT ON 11/20/19 ORIGINATED BY (B)(6), TECHNICAL QUALITY LEADER AT (B)(6) STATING THAT THEY RECEIVED A SINGLE REPORT FROM THEIR CUSTOMER, (B)(6), OF SEVEN INSTANCES OF PATIENT INFECTION FOLLOWING THE USE OF THE AVID MEDICAL ITEM NO. 140-01 BAL CATH PREP PK W/SPEC TRAP. THE DEVICE HOWEVER WAS USED IN CONJUNCTION WITH A SECOND MEDICAL CONVENIENCE KIT, WHICH IS NOT MANUFACTURED BY (B)(6). IT WAS REPORTED BY AVANOS MEDICAL THAT EACH PATIENT DEVELOPED THE SAME TYPE OF INFECTION FOLLOWING THE PROCEDURES. THE COMPLAINT LOT NUMBERS WERE NOT PROVIDED BY THE HOSPITAL AND THE DEVICES WERE NOT AVAILABLE FOR EVALUATION. (B)(6) ISSUED FORMAL COMPLAINT # (B)(4) FOR THE REPORTED EVENTS. AVANOS MEDICAL SUBMITTED THE FOLLOWING MDRS FOR THE SEVEN EVENTS: 3011270181-2019-00032, 3011270181-2019-00033, 3011270181-2019-00034, 3011270181-2019-00035, 3011270181-2019-00036, 3011270181-2019-00037 AND 3011270181-2019-00038. THERE HAVE BEEN NO STERILE LOAD FAILURES AT (B)(6). EACH BATCH OF THIS PRODUCT CODE MET A STERILITY ASSURANCE LEVEL OF 10-6 ON ALL LOADS MANUFACTURED N THE LAST 3 YEARS, WITH NO BIOLOGICAL INDICATOR FAILURES. ONE ITEM, VIAL, SODIUM CHLORIDE. 9,50ML (NDC 0409-4888-50), IS PROVIDED AS A PIGGYBACKED ITEM (PACKAGED EXTERNALLY NOT SUBJECT TO STERILIZATION BY (B)(6)) FOR THIS MEDICAL CONVENIENCE KIT. THIS SODIUM CHLORIDE PRODUCT WAS REPORTED TO PFIZER INC VIA PHONE BY (B)(6)'S COMPLAINT TEAM ON 11/25/19. AT THIS TIME THERE HAS BEEN NO RESPONSE FROM PFIZER TO DATE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1272269 AVID MEDICAL BAL CATH PREP PK W/SPEC TRAP LRO AVID MEDICAL, INC. A MEDICAL ACTION IND. COMPANY 140-10 UNKNOWN

Patients

Seq Age Sex Outcome Treatment
1 Other