FDA Adverse Event Other Summary report: N

COULTER AC T DIFF 2 ANALYZER

MDR report key: 947722 · Received November 12, 2007

Report

Report Number
1061932-2007-00054
Event Type
Other
Date Received
November 12, 2007
Date of Event
October 18, 2007
Report Date
November 12, 2007
Manufacturer
BECKMAN COULTER, INC.
Product Code
GKZ
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

QC IS RUN ONCE A DAY. QC WAS PERFORMED BEFORE AND AFTER THE EVENT AND RESULTS WERE IN RANGE. THE SPECIMEN WAS COLLECTED IN A BD, 4ML TUBE AND RAN LESS THAN 15 MINUTES AFTER COLLECTION. THE SAMPLE WAS TESTED IN CLOSED VIAL WHOLE BLOOD MODE. A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER'S LAB: A) PER FSE, QC WAS PERFORMING PROPERLY. B) THE FSE INSPECTED FLUIDICS AND VERIFIED MIXING BUBBLES. C) THE FSE PERFORMED REPRODUCIBILITY USING NORMAL CONTROL MATERIAL AND OBTAINED PASSING RESULTS. D) THE FSE VERIFIED THE INSTRUMENT'S PERFORMANCE. A CLEAR ROOT CAUSE HAS NOT BEEN DETERMINED IN THIS EVENT. A MALFUNCTION WILL BE ASSUMED FOR THE PURPOSE OF THIS REPORT.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) REGARDING LOW HEMOGLOBIN (HGB) AND PLATELET (PLT) RESULTS THAT WERE GENERATED BY THE COULTER AC T DIFF 2 INSTRUMENT FOR A SINGLE PATIENT SAMPLE. A PATIENT SAMPLE WAS TESTED FOR HGB AND PLT AND RESULTS OF 4.2G/DL AND 240 X 10 TO THE THIRD POWER CELLS/UL WERE OBTAINED RESPECTIVELY. THE RESULTS WERE REPORTED OUT OF THE LAB AND THE PATIENT WAS SENT TO A HOSPITAL AND REDRAWN. THE HOSPITAL'S RESULTS WERE HIGHER FOR BOTH ANALYTES; HGB RESULT WAS 9.1G/DL. PLT RESULT WAS 416 X 10 TO THE THIRD POWER CELLS/UL. THE HOSPITAL'S RESULTS WERE CONSIDERED CORRECT. ON THE NEXT DAY, THE CUSTOMER RE-TESTED THE ORIGINAL SAMPLE FOR HGB AND PLT AND RESULTS WERE: HGB RESULT WAS 9.7G/DL. PLT RESULT WAS 385 X 10 TO THE THIRD POWER CELLS/UL. BASED ON INFORMATION PROVIDED, THERE WAS NO AFFECT TO PATIENT OR USER AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COULTER AC T DIFF 2 ANALYZER AUTOMATED DIFFERENTIAL CELL COUNTER GKZ BECKMAN COULTER, INC. AC T DIFF 2 NA

Patients

Seq Age Sex Outcome Treatment
1 NA YR