Narratives

Additional Manufacturer Narrative · 1

THE RESULTS/METHOD AND CONCLUSION CODES ALONG WITH INVESTIGATION RESULTS WILL BE PROVIDED IN THE FINAL REPORT.

Additional Manufacturer Narrative · 1

AS RECEIVED, A COMPLETE LEAD WAS RETURNED IN ONE PIECE WITH THE HELIX FULLY EXTENDED AND CLOGGED WITH BLOOD/TISSUE. ELECTRICAL TESTING DID NOT FIND ANY INDICATION OF CONDUCTOR FRACTURES OR INTERNAL SHORTS. VISUAL INSPECTION AND X-RAY EXAMINATION OF THE LEAD DID NOT FIND ANY ANOMALIES.

Description of Event or Problem · 1

RELATED MANUFACTURER REFERENCE NUMBER:2017865-2019-17984, 2017865-2019-17985. IT WAS REPORTED THAT THE PATIENT PRESENTED FOR INITIAL IMPLANT. THE RIGHT VENTRICULAR LEAD (RV), RIGHT ATRIAL (RA) LEAD, AND LEFT VENTRICULAR (LV) LEAD WERE PLACED AND RESTED APPROPRIATELY. DURING PROCEDURE, FLUOROSCOPY TESTING TO CHECK FINAL LEAD PLACEMENT NOTED THE RV LEAD WAS NO LONGER IN ITS ORIGINAL POSITION. RIGHT VENTRICULAR PERFORATION WAS NOTED. THE PATIENT SUBSEQUENTLY DEVELOPED TACHYCARDIA AND HYPOTENSION. THE PHYSICIAN SUSPECTED TAMPONADE, AND TAMPONADE WAS CONFIRMED WITH INTRAOPERATIVE ECHOCARDIOGRAM. FURTHERMORE, NO PULSE WAS DETECTED AND ACLS PROTOCOL WAS IMITATED. PERICARDIOCENTESIS WAS PERFORMED AND TAMPONADE WAS RELIEVED. THE PATIENT¿S VITAL SIGNS WERE STABILIZED. THE ORIGIN OF THE TAMPONADE WAS NOT ENTIRELY KNOWN. THE PHYSICIAN FELT THE RV LEAD PERFORATED THE RIGHT VENTRICLE BUT DID NOT FELT IT WAS A RESULT OF THE LEAD ITSELF BUT POSSIBLY THAT THE RV LEAD COULD HAVE BEEN ADVANCED DURING PLACEMENT OF THE LV LEAD. ALL THE LEADS WERE REMOVED. THE PATIENT WAS STABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1265555 TENDRIL STS PERMANENT PACEMAKER ELECTRODE NVN ST. JUDE MEDICAL, INC.(CRM-SYLMAR) 2088TC/46 P000089367 05414734502870

Patients

Seq Age Sex Outcome Treatment
1 65 YR Required Intervention DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR| QUARTET