FDA Adverse Event Malfunction Summary report: N

PHILIPS MEDICAL SYSTEMS MX8000IDT SCANNER

MDR report key: 947048 · Received November 8, 2007

Report

Report Number
1525965-2007-00030
Event Type
Malfunction
Date Received
November 8, 2007
Date of Event
April 25, 2007
Report Date
April 25, 2007
Manufacturer
PHILIPS MEDICAL SYSTEMS
Product Code
JAK
PMA / PMN Number
K012009
Removal / Correction Number
1525962-09/04/07-004-C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE TEAL HAS UL ELECTRICAL SAFETY APPROVAL, AND THE RISK OF INJURY TO A PATIENT IS REMOTE. THIS ISSUE WAS FOUND DURING THE REVIEW OF INTERNAL SERVICE RECORDS. VOLUNTARY MEDICAL DEVICE RECALL REPORT WAS FILED. IT IS A CLASS 2 RECALL. PMS ISSUED A FIELD CHANGE ORDER TO CORRECT THIS ISSUE IN MAY, 2007. THIS FCO HAS BEEN CHANGED TO MANDATORY ACTION TO ENSURE AGREEMENT WITH THE C&R REPORT. THE ROOT CAUSE OF THE ISSUE IS A COMPONENT FAILURE ON THE VENDED DEVICE.

Description of Event or Problem · 1

BASED ON INFORMATION THAT PMS RECEIVED ON 10/11/2007, PMS HAS DETERMINED THAT THIS REPORT IS SIMILAR TO OTHER MDR REPORTABLE EVENTS INVOLVING THE TEAL. HERE IS A SUMMARY OF THE ISSUE. THE USER REPORTED AN: ELECTRIC SMELL NEAR THE CONSOLE. GANTRY IS SPARKING. THE INITIAL REPORT DID NOT DOCUMENT THAT THE TEAL PDU WAS CAUSING THE PROBLEM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PHILIPS MEDICAL SYSTEMS MX8000IDT SCANNER COMPUTER TOMOGRAPHY SCANNER JAK PHILIPS MEDICAL SYSTEMS

Patients

Seq Age Sex Outcome Treatment
1 YR