FDA Adverse Event Malfunction Summary report: N

PALMAZ MEDIUM ON POWERFLEX PLUS

MDR report key: 946991 · Received November 7, 2007

Doing adverse event research by hand?

The Research Assistant searches MAUDE for you, groups reports by failure mode, and cites every record.

Free to try · every answer cites its records

Report

Report Number
9610978-2007-00401
Event Type
Malfunction
Date Received
November 7, 2007
Date of Event
October 19, 2007
Report Date
October 19, 2007
Manufacturer
CORDIS EUROPA, N.V.
Product Code
FGE
PMA / PMN Number
NA
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

IT WAS REPORTED THAT AFTER THE PHYSICIAN REMOVED THE STENT DELIVERY SYSTEM FROM THE PACKAGE, HE TRIED TO RE-POSITION THE STENT ON THE BALLOON, HOWEVER, THE STENT "BROKE'. THERE WERE NO ANOMALIES NOTED WITH THE PRODUCT BEFORE REPOSITIONING THE STENT AND THERE WERE NO NOTED DEFECTS WITH THE PACKAGING. THIS COMPLAINT IS BEING REPORTED BECAUSE HAD THE MALFUNCTION OCCURRED INSIDE THE PATIENT, IT WOULD BE A REPORTABLE MALFUNCTION. BASED ON THE INFORMATION AVAILABLE, IT APPEARS THAT THE DIFFICULTY EXPERIENCED BY THE CUSTOMER MAY HAVE BEEN CAUSED BY USER HANDLING AND IS NOT RELATED TO A PRODUCT QUALITY ISSUE. A DEVICE HISTORY RECORD (DHR) REVIEW WAS PERFORMED AND SHOWED THAT THIS LOT OF PRODUCTS MET ALL REQUIREMENTS PER THE APPLICABLE MANUFACTURING QUALITY PLAN (MQP). THIS REVIEW WAS EXTENDED TO SUBASSEMBLY PART NUMBERS WITH LOT NUMBER 0311060241, 0410060519 AND 0410060598. THIS REVIEW CONFIRMED THAT SUBASSEMBLIES MET ALL REQUIREMENTS PER THE APPLICABLE MQP AS WELL.

Description of Event or Problem · 1

AFTER OPENED THE STERILE PACKAGE, THE PHYSICIAN TRIED TO RE-MOUNT THE STENT OF THE PRODUCT (PALMAZ), BUT THE STENT BROKE. ANOTHER STENT (PALMAZ, SIZE: UNKNOWN) WAS USED FOR THE PROCEDURE AND THE PROCEDURES WAS SUCCESSFULLY COMPLETED. THERE WERE NO ANOMALIES NOTED WITH THE PRODUCT BEFORE REPOSITIONING THE STENT. THERE WERE NO NOTED DEFECTS WITH THE PACKAGING. THE PRODUCT WAS NOT CLINICALLY; THEREFORE, THERE WAS NO ADVERSE CONSEQUENCES TO THE PATIENT. THE PRODUCT WILL NOT BE RETURNED FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PALMAZ MEDIUM ON POWERFLEX PLUS ENDOVASCULAR SDS/STENTS (FGE) FGE CORDIS EUROPA, N.V. NA R1106094

Patients

Seq Age Sex Outcome Treatment
1 49 YR