HEARTMATE MOBILE POWER UNIT, NA
Report
- Report Number
- 2916596-2019-05868
- Event Type
- Malfunction
- Date Received
- December 13, 2019
- Date of Event
- November 29, 2019
- Report Date
- March 17, 2020
- Manufacturer
- THORATEC CORPORATION
- Product Code
- DSQ
- UDI-DI
- 813024010883
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MD, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
SECTION A2: CORRECTED INFORMATION SECTION H4: ADDITIONAL INFORMATION MANUFACTURER'S INVESTIGATION CONCLUSION: THE REPORTED EVENT OF NO EXTERNAL POWER ALARMS WAS CONFIRMED. THE MPU WAS NOT RETURNED FOR ANALYSIS; HOWEVER, A LOG FILE WAS SUBMITTED. THE SUBMITTED LOG FILES CONTAINED APPROXIMATELY 10+ DAYS OF DATA FROM (B)(6)2019 TO (B)(6)2019, PER THE TIMESTAMP. NO EXTERNAL POWER ALARMS ASSOCIATED WITH A LOSS OF AC POWER WHILE THE CONTROLLER WAS CONNECTED TO A MPU WERE RECORDED ON (B)(6)2019 AND (B)(6)2019. NO OTHER ATYPICAL ALARMS WERE RECORDED IN THE REMAINING EVENTS AND THE CONTROLLER OPERATED THE PUMP AT THE SET SPEED THROUGHOUT THE LOG FILE. MULTIPLE ATTEMPTS WERE MADE TO OBTAIN ADDITIONAL INFORMATION FROM THE CUSTOMER REGARDING THE EVENT; HOWEVER, NO ADDITIONAL INFORMATION WAS PROVIDED. THE CAUSE OF THE REPORTED EVENT WAS UNABLE TO BE CONCLUSIVELY DETERMINED. REVIEW OF THE DEVICE HISTORY RECORDS FOUND THAT (B)(6) WAS MANUFACTURED WITH THE V-LOCK AC CONNECTOR. NO FURTHER INFORMATION WAS PROVIDED. THE MANUFACTURER IS CLOSING THE FILE ON THIS EVENT.
PMA/510(K) #: P060040; P160054. NO FURTHER INFORMATION WAS PROVIDED. THE MANUFACTURER IS CLOSING THE FILE ON THIS EVENT.
IT WAS REPORTED THAT THE CUSTOMER REQUESTED A LOG FILE REVIEW. THE MANUFACTURER TECHNICAL SERVICES REVIEWED THE LOG FILE AND OBSERVED LOW VOLTAGE HAZARD AND NO EXTERNAL POWER EVENTS ON 11/17/ AND 11/24, WHICH ARE DUE TO A TOTAL LOSS OF POWER FROM THE MOBILE POWER UNIT. NO FURTHER INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1260779 | HEARTMATE MOBILE POWER UNIT, NA | VENTRICULAR (ASSISST) BYPASS | DSQ | THORATEC CORPORATION | 107754 | 813024010883 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 40 YR | Hospitalization |