COMBUR 10 UX 100
Report
- Report Number
- 1823260-2019-04412
- Event Type
- Malfunction
- Date Received
- December 13, 2019
- Date of Event
- November 22, 2019
- Report Date
- July 29, 2020
- Manufacturer
- ROCHE DIAGNOSTICS
- Product Code
- JIL
- PMA / PMN Number
- K032437
- Removal / Correction Number
- RES81177
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SF
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THE CUSTOMER DID NOT RETURN ANY PRODUCT. THE SUSPECT MEDICAL DEVICE HAS BEEN UPDATED. APPLICABLE FIELDS OF SECTIONS D AND G WERE UPDATED. THE (B)(6) ANALYZER SERIAL NUMBER WAS (B)(6). SECTION H9 WAS UPDATED. DURING INTERNAL TESTING, ROCHE DETERMINED THE LIMITS OF DETECTION (LOD) FOR PROTEIN, NITRITE, LEUKOCYTES, AND ERYTHROCYTES ON THE (B)(6) URINE ANALYZER WITH CHEMSTRIP® 5 OB, CHEMSTRIP 7, CHEMSTRIP 10 MD, AND CHEMSTRIP 10 UA TEST STRIPS WERE HIGHER THAN WHAT IS LISTED IN THEIR RESPECTIVE TEST STRIP METHOD SHEETS. THIS CAN LEAD TO FALSE NEGATIVE RESULTS ON THE (B)(6) URINE ANALYZER FOR THE FOUR AFFECTED PARAMETERS. ROCHE HAS PROVIDED CUSTOMERS WITH THE FOLLOWING INSTRUCTIONS AND WORK AROUNDS: CHEMSTRIP 5 OB AND CHEMSTRIP 7 TEST STRIPS SHOULD ONLY BE READ VISUALLY. THEY CAN NO LONGER BE READ ON THE (B)(6)URINE ANALYZER. CHEMSTRIP 10 MD OR CHEMSTRIP 10 UA TEST STRIPS CAN BE USED WITH THE (B)(6) URINE ANALYZER; HOWEVER, A NEGATIVE RESULT FOR ANY ONE OF THE FOUR AFFECTED PARAMETERS (I.E., PROTEIN, NITRITE, LEUKOCYTES, AND ERYTHROCYTES) ON THE (B)(6) URINE ANALYZER MUST BE REPEATED WITH A NEW TEST STRIP THAT IS READ VISUALLY.
THIS EVENT OCCURRED IN (B)(6). THE ANALYZER HAS BEEN USED SINCE THIS EVENT AND ALL RESULTS HAVE BEEN ACCURATE. QC HAD PASSED.
THE INITIAL REPORTER COMPLAINED OF DISCREPANT RESULTS FOR 1 PATIENT TESTED FOR LEUKOCYTES ON A URISYS 1100 ANALYZER. THE RESULT FROM THE INSTRUMENT WAS NEGATIVE. WHEN THEY CHECKED VISUALLY THE RESULT WAS POSITIVE (500 LEUKOCYTES). THE COMBUR 10 TEST STRIP LOT NUMBER WAS 38055304 WITH AN EXPIRATION DATE OF 31-MAR-2020.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1264090 | COMBUR 10 UX 100 | URINE TEST STRIPS | JIL | ROCHE DIAGNOSTICS | U1100 | 38055304 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |