STYLE 68 SALINE FILLED BREAST IMPLANT
Report
- Report Number
- 9617229-2019-16037
- Event Type
- Injury
- Date Received
- December 13, 2019
- Date of Event
- August 15, 2019
- Report Date
- September 10, 2020
- Manufacturer
- ALLERGAN (COSTA RICA)
- Product Code
- FWM
- PMA / PMN Number
- P990074
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KS
- Reporter Occupation
- PHYSICIAN
Narratives
DEVICE EVALUATION: VISUAL ANALYSIS OF THE IDENTIFIED PHOTOS: DEVICE PATCH WITH LOT NUMBER 2100101, CREASE FOLD, WEAR ABRASION, AND APPARENTLY THE UNIT HAS AN OPENING AND LABELED AS RIGHT SIDE. DEVICE ANALYSIS PERFORMED THROUGH PHOTOGRAPHS, DUE TO THE IMPOSSIBILITY TO PERFORM MICROSCOPIC ANALYSIS IT IS NOT POSSIBLE TO DETERMINE THE MOST LIKELY FAILURE MODE. A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCES NOTED.
HEALTHCARE PROFESSIONAL REPORTED A RIGHT SIDE. DEVICE HAS BEEN EXPLANTED.
ALLERGAN DID NOT SUBMIT THIS MDR WITHIN 30 DAYS OF BECOMING AWARE. RECENT STIMULATED REPORTING RELATED TO 2011068-7/2/19-001-R HAS INCREASED COMPLAINT AND MDR VOLUME. ALLERGAN IS IMPLEMENTING A PLAN TO ADDRESS THE INCREASED VOLUMES. REASON FOR REOPERATION DUE TO DEFLATION. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. THIS IS A KNOWN POTENTIAL ADVERSE EVENT ADDRESSED IN THE PRODUCT LABELING.
HEALTHCARE PROFESSIONAL REPORTED A DEFLATION ON AN UNKNOWN SIDE. DEVICE REMAINS IMPLANTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1260745 | STYLE 68 SALINE FILLED BREAST IMPLANT | PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE | FWM | ALLERGAN (COSTA RICA) | 2100101 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 32 YR | Required Intervention |