FDA Adverse Event Malfunction Summary report: N

IPC® HANDPIECE - STYLUS TOUCH®

MDR report key: 9450893 · Received December 11, 2019

Report

Report Number
1045254-2019-00666
Event Type
Malfunction
Date Received
December 11, 2019
Date of Event
November 4, 2019
Report Date
January 28, 2020
Manufacturer
MEDTRONIC XOMED INC.
Product Code
HBE
UDI-DI
00643169945371
PMA / PMN Number
K081475
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

PREVIOUS PMA/510K: K050798 NO LONGER APPLIES. MEDTRONIC IS SUBMITTING THIS REPORT TO COMPLY WITH FDA REPORTING REGULATIONS UNDER 21 CFR PARTS 4 AND 803. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION AND HAS PROVIDED AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY ¿DEFECTS¿ OR HAS ¿MALFUNCTIONED¿. THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REGULATORY REPORTING. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE THEM BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

ANALYSIS FOUND NO FAULT WITH THE DEVICE. PER REPAIR NOTES: THE HAND PIECE WAS TESTED AND THE AMBIENT TEMPERATURE OR START TEMPERATURE WAS 76.9F AND THE HIGHEST OR MAXIMUM TEMPERATURE WAS 102.8F (WHICH WAS PASSING). THE MAXIMUM RISE OF TEMPERATURE FOR THE DRILL WAS 27F. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

A HEALTHCARE PROVIDER (HCP) REPORTED THAT THE HAND PIECE HAD EXCESSIVE HEAT. THERE WAS NO PATIENT IMPACT. THE EVENT OCCURRED DURING SERVICING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1246526 IPC® HANDPIECE - STYLUS TOUCH® DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) HBE MEDTRONIC XOMED INC. EM210RF 209914503 00643169945371

Patients

Seq Age Sex Outcome Treatment
1