FDA Adverse Event Malfunction Summary report: N

BLADDER CATHETER, PVC, NELATON-ROBINSON

MDR report key: 9446987 · Received December 10, 2019

Report

Report Number
9610520-2019-00026
Event Type
Malfunction
Date Received
December 10, 2019
Date of Event
October 27, 2019
Report Date
November 12, 2019
Manufacturer
TELEFLEX MEDICAL
Product Code
KOD
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
RISK MANAGER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE COMPLAINT IS REGARDING 6 BLADDER CATHETERS, NELATON-ROBINSON WITH P/N 220500 WHERE CUSTOMER REPORTED THAT CATHETER SHAFT LEAKED - DEVICES WERE TESTED PRIOR TO USE AND ALL 6 CATHETERS SHOWED A LEAKAGE. THE ACTUAL 6 DEVICES WERE RETURNED FOR INVESTIGATION. A LEAK TEST ACCORDING TO THE EFFECTIVE WORK INSTRUCTION FOR THE TIGHTNESS TEST BETWEEN FUNNEL AND TUBE WAS PERFORMED ON ALL 6 SAMPLES. THE FUNCTIONAL TEST CONFIRMED A LEAK IN EACH OF THE 6 SAMPLES. THE LEAK OCCURS BETWEEN THE CATHETER TUBE AND THE FUNNEL. IN ADDITION, A REVIEW FOR THE PRODUCTION DATA FOR LOT 18/28/212 WAS PERFORMED. 6,550 CATHETERS WERE GLUED BY THE FUNNEL-GLUING-MACHINE NO. 4. THE PERFORMED IN-PROCESS CONTROL WHICH ALSO INCLUDED A LEAK TEST REVEALED NO DEVIATIONS DURING THE PRODUCTION OF THE CLAIMED LOT. BASED ON THE INVESTIGATION, THE SUPPLIER CONFIRMED THE REPORTED ISSUE FOR THE 6 ACTUAL SAMPLES. A NC HAS BEEN OPENED BY THE SUPPLIER, BUT NOT CLEAR CAUSE FOR THIS FAILURE COULD BE DETERMINED. THE REPORTED FAILURE BETWEEN FUNNEL AND CATHETER TUBE OCCURS SPORADICALLY AND THIS IS AN INCONVENIENCE FOR THE USER BUT NO RISK FOR THE USER. DURING THE IN-PROCESS CONTROL, THE CONNECTION IS TESTED AT 1 BAR PRESSURE AND NO DEVIATION WAS REPORTED FOR THIS LOT. TELEFLEX WILL CONTINUE TO MONITOR AND TREND RELATED EVENTS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CATHETER SHAFT LEAKED. DEVICES WERE TESTED PRIOR TO USE AND ALL 6 CATHETERS SHOWED A LEAKAGE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1243463 BLADDER CATHETER, PVC, NELATON-ROBINSON KOD TELEFLEX MEDICAL 18/28/212

Patients

Seq Age Sex Outcome Treatment