FDA Adverse Event Malfunction Summary report: N

TRANSPAC® IV MONITORING KIT WITH SAFESET¿ RESERVOIR AND 2 BLOOD SAMPLING PORTS,

MDR report key: 9439981 · Received December 9, 2019

Report

Report Number
9617594-2019-00417
Event Type
Malfunction
Date Received
December 9, 2019
Date of Event
November 13, 2019
Report Date
November 20, 2019
Manufacturer
ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V.
Product Code
DRS
UDI-DI
00840619041199
PMA / PMN Number
K061573
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION D10: THE DEVICE WAS RETURNED FOR INVESTIGATION ON (B)(6)2019 FOR INVESTIGATION. H10: RECEIVED ONE UNUSED LIST# 011-46103-87, TRANSPAC® IV MONITORING KIT WITH SAFESET¿ RESERVOIR AND 2 BLOOD SAMPLING PORTS, 84" TUBING, DISPOSABLE TRANSDUCER, 03 ML SQUEEZE FLUSH, MACRODRIP (POLE MOUNT). THE REPORTED COMPLAINT OF "DISCONNECTED LINE" WAS CONFIRMED. DURING VISUAL INSPECTION, THE TUBING BREAKAGE WAS OBSERVED AT THE SAFESET POCKET. A PULL TEST WAS CONDUCTED AT A REPRESENTATIVE BOND AND DID NOT FIND ANYTHING UNUSUAL. THE PROBABLE CAUSE FOR THE DISCONNECTED LINE COULD NOT BE DETERMINED. A DEVICE HISTORY REVIEW (DHR) LOT# 4102146 AND RELEVANT COMMODITIES WERE REVIEWED, AND NO NON-CONFORMANCES WERE FOUND THAT WOULD HAVE CONTRIBUTED TO THE REPORTED COMPLAINT.

Additional Manufacturer Narrative · 1

THE DEVICE IS AVAILABLE FOR INVESTIGATION. IT HAS NOT BEEN RECEIVED.

Description of Event or Problem · 1

THE EVENT INVOLVED A TRANSPAC IV MONITORING KIT WITH SAFESET RESERVOIR AND 2 BLOOD SAMPLING PORTS 84" TUBING DISPOSABLE TRANSDUCER, 03 ML SQUEEZE FLUSH, MACRODRIP (POLE MOUNT) THAT THE CUSTOMER REPORTED THE LINE DISCONNECTED. THE CUSTOMER REPORTED THE ANESTHESIA TECHNOLOGIST NOTICED THE LINE DISCONNECTED WHERE IT SHOULD BE BONDED TOGETHER. THERE WAS PATIENT INVOLVEMENT, HOWEVER, NO REPORT OF ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1233857 TRANSPAC® IV MONITORING KIT WITH SAFESET¿ RESERVOIR AND 2 BLOOD SAMPLING PORTS, TRANSDUCER, BLOOD-PRESSURE, EXTRAVASCULAR DRS ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V. 4102146 00840619041199

Patients

Seq Age Sex Outcome Treatment
1