FDA Adverse Event Malfunction Summary report: N

ADMINISTRATION SET FOR HALO II AMBULATORY INFUSION PUMP

MDR report key: 9437210 · Received December 9, 2019

Report

Report Number
3011581906-2019-00045
Event Type
Malfunction
Date Received
December 9, 2019
Date of Event
November 15, 2019
Report Date
November 25, 2019
Manufacturer
INFUTRONIX, LLC
Product Code
FPA
UDI-DI
00817170020031
PMA / PMN Number
K153193
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

IMAGES OF THE AFFECTED ADMINISTRATION SET AS IT WAS RECEIVED BY (B)(4) WERE PROVIDED ON 11/27/2019. THE LOCATION WHERE THE TUBING CONNECTS TO THE CASSETTE HAD BEEN BROKEN AND DETACHED. THE LOT NUMBER OF THE AFFECTED ADMINISTRATION SET WAS UNKNOWN. HOWEVER, BASED OFF THE IMAGES OF THE ADMINISTRATION SET PROVIDED BY THE DISTRIBUTOR, THIS ADMINISTRATION SET WAS MANUFACTURED WITH THE NEW DESIGN OF THE CASSETTE CONNECTION. THE DISTRIBUTOR CONFIRMED THAT THE ONLY ADMINISTRATION SETS WITH THE NEW DESIGN SENT TO THE END USER WERE FROM LOT # 1903007. FURTHER INVESTIGATION FOUND THAT THIS LOT # 1903007 WAS ALSO TESTED AS PART OF A SCAR BY THE CONTRACT MANUFACTURER. THIS LOT # 1903007 WAS TESTED TO TENSILE STRENGTH REQUIREMENTS PER ISO 8536-9(2015) AND PASSED ALL TESTING CRITERIA. THIS LEADS TO THE CONCLUSION THAT THE BROKEN ADMINISTRATION SET CONNECTION WAS DUE TO EXCESSIVE PHYSICAL FORCE.

Description of Event or Problem · 1

ON 11/25/2019, A DISTRIBUTOR OF INFUTRONIX REPORTED A TUBING LEAKING ISSUE: "STOPPED INFUSION ABOUT 8 HOURS INTO A 46 HOUR INFUSION. THE PT. STATED THAT SHE THOUGHT THE BAG WAS LEAKING SO SHE DOUBLE BAGGED EVERYTHING. WHEN SHE CAME IN THIS MORNING, THE BAG OF 5FU HAD LEAKED AND WAS COMPLETELY EMPTY. THE PLASTIC CONNECTOR BETWEEN THE TUBING ON THE OUTSIDE OF THE CASSETTE AND THE TUBING INSIDE OF THE CASSETTE HAD BROKEN." THE CONTRACT MANUFACTURER OF THE AFFECTED DEVICE IS PODO XINGDA (TIANJIN) MEDICAL CO. LTD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1236621 ADMINISTRATION SET FOR HALO II AMBULATORY INFUSION PUMP IV ADMINISTRATION SET FPA INFUTRONIX, LLC HS-004 1903007 00817170020031

Patients

Seq Age Sex Outcome Treatment
1 63 YR