FDA Adverse Event Malfunction Summary report: N

MTS INCUBATOR

MDR report key: 942952 · Received November 13, 2007

Report

Report Number
1056600-2007-00313
Event Type
Malfunction
Date Received
November 13, 2007
Date of Event
October 19, 2007
Report Date
November 13, 2007
Manufacturer
MICRO TYPING SYSTEMS, INC.
Product Code
JTQ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WY, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

USER ERROR IDENTIFIED AS ROOT CAUSE. SALINE LEAKED INSIDE INCUBATOR. PRODUCT LABELING STATES "DO NOT POUR LIQUID OVER THE INSTRUMENT." (MTS INCUBATOR USER'S GUIDE 2004-06-30 P.9) ASSISTANCE PROVIDED TO THE CUSTOMER OVER THE PHONE HAS RESOLVED THE ISSUE. INSTRUMENT HAS BEEN REMOVED FROM SERVICE AND DISPOSED BY CUSTOMER.

Description of Event or Problem · 1

THE CUSTOMER INDICATED THAT THE INCUBATOR GENERATED SMOKE. THE CUSTOMER HAD A 3 GALLON CONTAINER OF SALINE ABOVE THE INCUBATOR. THE SALINE LEAKED INSIDE THE INCUBATOR AND MELTED THE CIRCUIT BOARD INSIDE CAUSING SMOKE TO GENERATE FROM THE INSIDE. NO FIRE OR SHOCK REPORTED AS A RESULT OF THIS INCIDENT. NO DEATH OR SERIOUS INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MTS INCUBATOR MTS GEL CARD INCUBATOR JTQ MICRO TYPING SYSTEMS, INC. NA NA

Patients

Seq Age Sex Outcome Treatment
1 YR