FDA Adverse Event
Malfunction
Summary report: N
MTS INCUBATOR
MDR report key: 942952
·
Received November 13, 2007
Report
- Report Number
- 1056600-2007-00313
- Event Type
- Malfunction
- Date Received
- November 13, 2007
- Date of Event
- October 19, 2007
- Report Date
- November 13, 2007
- Manufacturer
- MICRO TYPING SYSTEMS, INC.
- Product Code
- JTQ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WY, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
USER ERROR IDENTIFIED AS ROOT CAUSE. SALINE LEAKED INSIDE INCUBATOR. PRODUCT LABELING STATES "DO NOT POUR LIQUID OVER THE INSTRUMENT." (MTS INCUBATOR USER'S GUIDE 2004-06-30 P.9) ASSISTANCE PROVIDED TO THE CUSTOMER OVER THE PHONE HAS RESOLVED THE ISSUE. INSTRUMENT HAS BEEN REMOVED FROM SERVICE AND DISPOSED BY CUSTOMER.
Description of Event or Problem · 1
THE CUSTOMER INDICATED THAT THE INCUBATOR GENERATED SMOKE. THE CUSTOMER HAD A 3 GALLON CONTAINER OF SALINE ABOVE THE INCUBATOR. THE SALINE LEAKED INSIDE THE INCUBATOR AND MELTED THE CIRCUIT BOARD INSIDE CAUSING SMOKE TO GENERATE FROM THE INSIDE. NO FIRE OR SHOCK REPORTED AS A RESULT OF THIS INCIDENT. NO DEATH OR SERIOUS INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MTS INCUBATOR | MTS GEL CARD INCUBATOR | JTQ | MICRO TYPING SYSTEMS, INC. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | YR |