FDA Adverse Event Injury Summary report: N

STYLE 68 SALINE FILLED BREAST IMPLANT

MDR report key: 9424258 · Received December 6, 2019

Report

Report Number
9617229-2019-19501
Event Type
Injury
Date Received
December 6, 2019
Date of Event
December 13, 2018
Report Date
April 1, 2020
Manufacturer
ALLERGAN (COSTA RICA)
Product Code
FWM
UDI-DI
10888628000193
PMA / PMN Number
P990074
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

DEVICE EVALUATION: THE DEVICE RELATED TO THE REPORTED EVENT OF DEFLATION, RECEIVED ON MAR 04, 2020 WITH LOT NUMBER 2930363. VISUAL ANALYSIS OF THE RETURNED DEVICE IDENTIFIED: CREASE FOLD, WEAR ABRASION, YELLOW PARTICLES IN THE SHELL, BROWN PARTICLES IN THE FILL CHANNEL AND ONE OPENING LINEAR ON THE POSTERIOR. LEAK TEST AND MICROSCOPIC ANALYSIS WAS PERFORMED WHICH IDENTIFIED: ONE OPENING CREASE SHARP ON THE POSTERIOR. THE FILL TEST INSPECTION WAS PERFORMED, THE RESULT IS NO BLOCKAGE. BASED ON THE DEVICE ANALYSIS THE FINAL ASSESSMENT IS: - ONE CREASE SHARP OPENING ON THE POSTERIOR ASSESSED AS FOLD FLAW OPENING. ADDITIONAL, CHANGED, AND/OR CORRECTED DATA: B.5, D.7, D.10, G.1, H.3, H.6.

Description of Event or Problem · 0

THE DEVICE HAS BEEN EXPLANTED.

Additional Manufacturer Narrative · 1

A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN INITIATED. IF ANY NEW, CHANGED OR CORRECTED INFORMATION IS NOTED, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. REASON FOR REOPERATION: WILL BE DEFLATION. THIS IS A KNOWN POTENTIAL ADVERSE EVENT ADDRESSED IN THE PRODUCT LABELING.

Description of Event or Problem · 1

HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE DEFLATION. THE DEVICE REMAINS IMPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1220040 STYLE 68 SALINE FILLED BREAST IMPLANT PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE FWM ALLERGAN (COSTA RICA) 68-240 2930363 10888628000193

Patients

Seq Age Sex Outcome Treatment
1 68 YR Required Intervention