FDA Adverse Event Malfunction Summary report: N

20G X 1.16IN (1.1 X 30 MM) W/ Y INTIMA II

MDR report key: 9419070 · Received December 5, 2019

Report

Report Number
3006948883-2019-01034
Event Type
Malfunction
Date Received
December 5, 2019
Date of Event
January 9, 2019
Report Date
January 3, 2020
Manufacturer
BD (SUZHOU)
Product Code
FOZ
PMA / PMN Number
K143610
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION SUMMARY: A DEVICE HISTORY REVIEW WAS CONDUCTED FOR LOT NUMBER 7019420. OUR RECORDS SHOW THAT THIS IS THE ONLY INSTANCE OF THIS ISSUE OCCURRING IN THIS PRODUCTION BATCH. ACCORDING TO THE SAMPLING PLAN APPLIED FOR PRODUCT PERFORMANCE, THIS LOT WAS ACCEPTED AND RELEASED WITHOUT DEFECTS BEING NOTED DURING THE FINAL ASSEMBLY OR VISUAL INSPECTIONS. A SAMPLE COULD NOT BE OBTAINED FOR EVALUATION AND TESTING. TWO RETAINED SAMPLES WERE EVALUATED AND PASSED A LEAKAGE TEST. WITHOUT THE ABILITY TO INVESTIGATE THE AFFECTED UNIT OUR QUALITY ENGINEERS WERE UNABLE TO DETERMINE THE ROOT CAUSE FOR THIS COMPLAINT.

Description of Event or Problem · 0

IT WAS REPORTED THAT LEAKAGE OCCURRED DURING USE WITH A 20G X 1.16IN (1.1 X 30 MM) W/ Y INTIMA II. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, "INTRAVENOUS INJECTION WAS PERFORMED ON THE PATIENT AND INDWELLING NEEDLE LEAKAGE WAS FOUND."

Additional Manufacturer Narrative · 1

A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT LEAKAGE OCCURRED DURING USE WITH A 20G X 1.16IN (1.1 X 30 MM) W/ Y INTIMA II. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, "INTRAVENOUS INJECTION WAS PERFORMED ON THE PATIENT AND INDWELLING NEEDLE LEAKAGE WAS FOUND."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1215481 20G X 1.16IN (1.1 X 30 MM) W/ Y INTIMA II INTRAVASCULAR CATHETER FOZ BD (SUZHOU) 7019420

Patients

Seq Age Sex Outcome Treatment
1 Other