FDA Adverse Event Malfunction Summary report: N

5.0MM FLEXIBLE SHAFT 620MM

MDR report key: 9416817 · Received December 5, 2019

Report

Report Number
2939274-2019-62494
Event Type
Malfunction
Date Received
December 5, 2019
Report Date
November 15, 2019
Manufacturer
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
Product Code
HTO
UDI-DI
10886982194138
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. THE COMPLAINT IS CONFIRMED. A DEFINITIVE ROOT CAUSE COULD NOT BE DETERMINED, IT IS POSSIBLE THAT ROUGH HANDLING DURING USE AND/OR PROCESSING COULD HAVE CONTRIBUTED TO THE COMPLAINT CONDITION. DURING THE INVESTIGATION NO UNIDENTIFIED PRODUCT DESIGN/MANUFACTURING ISSUES OR DISCREPANCIES WERE OBSERVED THAT MAY HAVE CONTRIBUTED TO THE COMPLAINT CONDITION. BASED ON THE INVESTIGATION FINDINGS, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST MARKET SAFETY SURVEILLANCE ACTIVITIES. DEVICE HISTORY LOT. PART: 352.044-US. LOT: L930740. MANUFACTURING SITE: BETTLACH. RELEASE TO WAREHOUSE DATE: (B)(6) 2018. A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED DEVICE LOT NUMBER, AND NO NON-CONFORMANCES WERE IDENTIFIED. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 0

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 0

THIS IS 2 OF 5 FOR REPORT (B)(4).

Additional Manufacturer Narrative · 1

COMPLAINANT PART IS EXPECTED TO BE RETURNED FOR MANUFACTURER REVIEW/INVESTIGATION, BUT HAS YET TO BE RECEIVED. WITHOUT A LOT NUMBER THE DEVICE HISTORY RECORDS REVIEW COULD NOT BE COMPLETED. PRODUCT WAS NOT RETURNED. BASED ON THE INFORMATION AVAILABLE, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. THIS COMPLAINT WILL BE ACCOUNTED FOR AND MONITORED VIA POST MARKET SURVEILLANCE ACTIVITIES. IF ADDITIONAL INFORMATION IS MADE AVAILABLE, THE INVESTIGATION WILL BE UPDATED AS APPLICABLE. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

IT WAS REPORTED THAT ON (B)(6) 2019, DURING EQUIPMENT CHECK, ONE (1) FLEXIBLE HEXAGONAL SCREWDRIVER SNAPPED IN HALF DUE TO IMPROPER HANDLING, TWO (2) FLEXIBLE SHAFT BENT AND UN-USABLE, ONE (1) HELICAL BLADE INSERTER, IS CAUGHT IN WITH THE COUPLING SCREW AND THE INSERTER HANDLE AND WON¿T ADVANCE, ONE (1) HELICAL BLADE/SCREW COUPLING SCREW WILL NOT ADVANCE AND THEREFORE CANNOT ENGAGE THE BLADE, THEY WERE DISCOVERED DAMAGED IN THE STERILE PROCESSING DEPARTMENT. THERE WAS NO PATIENT INVOLVEMENT. THIS COMPLAINT INVOLVES FIVE (5) DEVICES. ONLY FOUR ARE REPORTABLE. THIS IS 2 OF 4 FOR REPORT (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1210694 5.0MM FLEXIBLE SHAFT 620MM REAMER HTO WRIGHTS LANE SYNTHES USA PRODUCTS LLC 352.044 10886982194138

Patients

Seq Age Sex Outcome Treatment
1