FDA Adverse Event Injury Summary report: N

CUP: MPACT ACETABULAR SHELL Ø54 TWO-HOLES

MDR report key: 9409888 · Received December 4, 2019

Report

Report Number
3005180920-2019-01030
Event Type
Injury
Date Received
December 4, 2019
Date of Event
November 11, 2019
Report Date
December 4, 2019
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
LPH
UDI-DI
07630030810824
PMA / PMN Number
K132879
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
IS
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 1

BATCH REVIEW PERFORMED ON 22 NOVEMBER 2019. LOT 182590: 144 ITEMS MANUFACTURED AND RELEASED ON 24-JUL-18. EXPIRATION DATE: 2023-07-10. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, 133 ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITH ANOTHER SIMILAR REPORTED EVENT. ADDITIONAL IMPLANT INVOLVED: STEM: MINIMAX 01.13.102R CEMENTLESS ANATOMICAL STEM RIGHT SIZE 2 (K170845) LOT. 181827. BATCH REVIEW PERFORMED ON 22 NOVEMBER 2019. LOT 181827: 15 ITEMS MANUFACTURED AND RELEASED ON 11-JUL-18. EXPIRATION DATE: 2023-07-03. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, 11 ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITH ANOTHER SIMILAR REPORTED EVENT (EVENT REGISTERED ON THE SAME LOT IS CAUSED BY TRAUMA). CLINICAL EVALUATION PERFORMED BY MEDICAL AFFAIRS MANAGER. HIP REVISION SURGERY PERFORMED 8 MONTHS AFTER CEMENTLESS TOTAL HIP ARTHROPLASTY IN A 70 YEAR OLD WOMAN. RADIOGRAPHIC IMAGES PROVIDED SHOW STEM TILTING FROM THE INITIAL POSITION PROBABLY DUE TO UNDER SIZING. THE REASON OF THIS STEM SIZE CHOICE IS UNKNOWN: PATIENT ANATOMY AND BONE MORPHOLOGY MAY HAVE CONTRIBUTED BUT THIS CANNOT BE CONFIRMED ON THE BASIS OF POSTOPERATIVE ANTEROPOSTERIOR IMAGES. WE CANNOT DETERMINE THE CAUSES OF THIS EVENT, BUT WE SEE NO REASON TO THINK THAT THE PROSTHETIC IMPLANTS WERE DEFECTIVE.

Description of Event or Problem · 1

REVISION SURGERY PERFORMED 9 MONTHS AFTER THE PRIMARY DUE TO LOOSE AND TILTED FEMORAL STEM AND MALPOSITIONED CUP. THE CUP AND THE STEM WERE REVISED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1205564 CUP: MPACT ACETABULAR SHELL Ø54 TWO-HOLES ACETABULAR CUP LPH MEDACTA INTERNATIONAL SA 182590 07630030810824

Patients

Seq Age Sex Outcome Treatment
1 70 YR Required Intervention