BD PEGASUS¿ SAFETY CLOSED IV CATHETER SYSTEM
Report
- Report Number
- 3006948883-2019-01006
- Event Type
- Malfunction
- Date Received
- December 3, 2019
- Date of Event
- November 14, 2019
- Report Date
- January 15, 2020
- Manufacturer
- BD (SUZHOU)
- Product Code
- FOZ
- PMA / PMN Number
- N/A
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- NURSE
Narratives
H.6. INVESTIGATION: A DEVICE HISTORY REVIEW WAS CONDUCTED FOR LOT NUMBER 9141631. OUR RECORDS SHOW THAT THIS IS THE ONLY INSTANCE OF THIS ISSUE OCCURRING IN THIS PRODUCTION BATCH. ACCORDING TO THE SAMPLING PLAN APPLIED FOR PRODUCT PERFORMANCE, THIS LOT WAS ACCEPTED AND RELEASED WITHOUT DEFECTS BEING NOTED DURING THE FINAL ASSEMBLY OR VISUAL INSPECTIONS. ADDITIONALLY, A SAMPLE WAS RECEIVED FOR THE PURPOSE OF OUR INVESTIGATION. OUR INVESTIGATORS NOTED A SMALL CRACK WAS FOUND DURING LEAKAGE TESTING, THAT WAS LOCATED ON THE ADAPTER. THE RETURNED PRODUCT'S SWAGE DIMENSIONS WERE MEASURED BY OUR TEAM AND FOUND TO BE WITHIN PRODUCT SPECIFICATIONS. UNFORTUNATELY, THE ROOT CAUSE FOR THIS COMPLAINT COULD NOT BE DETERMINED AT THE CONCLUSION OF OUR REVIEW. H3 OTHER TEXT: SEE H.10.
IT WAS REPORTED THAT THE BD PEGASUS¿ SAFETY CLOSED IV CATHETER SYSTEM NEEDLE HUB WAS BROKEN DURING USE AND CAUSED BLOOD LEAKAGE AFTER THE PENETRATION. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM CHINESE TO ENGLISH: "IT'S NOTICED THAT NEEDLE HUB BROKEN AND LEAD TO BLOOD LEAKAGE AT THE NEEDLE HUB AFTER COMPLETED PENETRATION."
A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED THAT THE BD PEGASUS¿ SAFETY CLOSED IV CATHETER SYSTEM NEEDLE HUB WAS BROKEN DURING USE AND CAUSED BLOOD LEAKAGE AFTER THE PENETRATION. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM (B)(6) TO ENGLISH: "IT'S NOTICED THAT NEEDLE HUB BROKEN AND LEAD TO BLOOD LEAKAGE AT THE NEEDLE HUB AFTER COMPLETED PENETRATION".
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1202071 | BD PEGASUS¿ SAFETY CLOSED IV CATHETER SYSTEM | INTRAVASCULAR CATHETER | FOZ | BD (SUZHOU) | 9141631 |