CART 9734056 S7 STAFF SHRT 100-120V INTL
Report
- Report Number
- 1723170-2019-05818
- Event Type
- Malfunction
- Date Received
- November 27, 2019
- Date of Event
- November 19, 2019
- Report Date
- January 30, 2020
- Manufacturer
- MEDTRONIC NAVIGATION, INC
- Product Code
- HAW
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- 003
Narratives
OTHER RELEVANT DEVICE(S) ARE: PRODUCT ID: 9734252, SERIAL/LOT #: UNKNOWN, UBD: UNKNOWN, UDI#: UNKNOWN. NO 510K AVAILABLE. NO PARTS HAVE BEEN RECEIVED BY THE MANUFACTURER FOR EVALUATION. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
A MEDTRONIC REPRESENTATIVE EVALUATED THE RETURNED ARM. IT WAS FOUND THAT THE STARBURST END OF THE ARM WAS STIFF AND WOULD NOT LOCK DOWN. THE HARDWARE INVESTIGATION FOUND THAT THE REPORTED EVENT WAS RELATED TO A HARDWARE ISSUE. THIS ISSUE WAS DOCUMENTED IN A MEDTRONIC NAVIGATION HARDWARE ANOMALY TRACKING DATABASE. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
OTHER RELEVANT DEVICE(S) ARE: PRODUCT ID: 960-537, SERIAL/LOT #: (B)(4). IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION SYSTEM. IT WAS REPORTED THAT A FIXATION FAILURE OF THE SUPPORT ARM WAS CONFIRMED. THE OBSOLETED SUPPORT ARM WAS DAMAGED AND COULD NOT BE REPAIRED. THERE WAS NO PATIENT PRESENT WHEN THIS ISSUE WAS IDENTIFIED.
ADDITIONAL INFORMATION WAS RECEIVED. IT WAS REPORTED THAT THE CAUSE OF THE FIXATION FAILURE WAS DUE TO DETERIORATION AND THAT THE ARM'S WIRE HAD DETERIORATED. IT WAS NOTED THAT PART REPLACEMENT RESOLVED THE ISSUE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1180481 | CART 9734056 S7 STAFF SHRT 100-120V INTL | INSTRUMENT, STEREOTAXIC | HAW | MEDTRONIC NAVIGATION, INC | 9734056 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |