Narratives
H.6 INVESTIGATION: DEVICE RETURNED WITH SPOTS ON PACKAGING THE DHR WERE REVIEWED FOR THIS BATCH AND THERE WAS NO RECORD OF NON-CONFORMANCE ASSOCIATED WITH THIS BATCH. BASED ON THE INVESTIGATION, THE ROOT CAUSE OF THIS COMPLAINT WAS CREATED DURING THE STEAM STERILIZATION PROCESS. H3 OTHER TEXT : SEE H.10
IT WAS REPORTED THAT THE BD POSIFLUSH¿ XS PRE-FILLED FLUSH SYRINGE NACL 0.9% THERE WERE ISSUES WITH FOREIGN MATTER. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM ITALIAN TO ENGLISH: WE HEREBY INFORM YOU THAT FOLLOWING IN-PROCESS CHECKS WE HAVE FOUND THE FOLLOWING ISSUES REGARDING THE FOLLOWING DELIVERIES OF COD. 306572 [REF. DRM: CODE BD SALINE XS]: LOT 9142652 - PCS PRE-CUT (34), PCS SPOTS ON PACKAGING (122), PCS SPOTS ON SYRINGE (12), PCS MISSING (2). LOT 9182101 - PCS PRE-CUT (16), PCS SPOTS ON PACKAGING (40), PCS CORPUSCLES WELDING (12). LOT 9199773 - PC PRE-CUT (10), PCS SPOTS ON PACKAGING (104), PZ. CORPUSCLES WELDING (7). LOT 9178438 - PC PRE-CUT (9), PCS SPOTS ON PACKAGING (18), PCS CORPUSCLES WELD (19), PCS STAINS ON SYRINGE (2), PCS MISSING (10). LOT 9213635 - PC PRE-CUT (30), PCS STAINS ON PACKAGING (34), PCS STAINS ON SYRINGE (1)
DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. THERE WERE MULTIPLE LOT NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION FOR EACH LOT NUMBER IS AS FOLLOWS: MEDICAL DEVICE LOT #: 9142652. MEDICAL DEVICE EXPIRATION DATE: 2022-04-30. DEVICE MANUFACTURE DATE: 2019-05-22. MEDICAL DEVICE LOT #: 9182101. MEDICAL DEVICE EXPIRATION DATE: 2022-06-30. DEVICE MANUFACTURE DATE: 2019-07-01. MEDICAL DEVICE LOT #: 9199773. MEDICAL DEVICE EXPIRATION DATE: 2022-06-30. DEVICE MANUFACTURE DATE: 2019-07-18. MEDICAL DEVICE LOT #: 9178438 . MEDICAL DEVICE EXPIRATION DATE: 2022-06-30. DEVICE MANUFACTURE DATE: 2019-06-27. MEDICAL DEVICE LOT #: 9213635. MEDICAL DEVICE EXPIRATION DATE: 2022-07-31. DEVICE MANUFACTURE DATE: 2019-08-01. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION AND/OR DEVICE HISTORY REVIEW, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED THAT THE BD POSIFLUSH¿ XS PRE-FILLED FLUSH SYRINGE NACL 0.9% THERE WERE ISSUES WITH FOREIGN MATTER. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM (B)(6) TO ENGLISH: WE HEREBY INFORM YOU THAT FOLLOWING IN-PROCESS CHECKS WE HAVE FOUND THE FOLLOWING ISSUES REGARDING THE FOLLOWING DELIVERIES OF COD. 306572: LOT 9142652 - PCS PRE-CUT (34), PCS SPOTS ON PACKAGING (122), PCS SPOTS ON SYRINGE (12), PCS MISSING (2). LOT 9182101 - PCS PRE-CUT (16), PCS SPOTS ON PACKAGING (40), PCS CORPUSCLES WELDING (12). LOT 9199773 - PC PRE-CUT (10), PCS SPOTS ON PACKAGING (104), PZ. CORPUSCLES WELDING (7). LOT 9178438 - PC PRE-CUT (9), PCS SPOTS ON PACKAGING (18), PCS CORPUSCLES WELD (19), PCS STAINS ON SYRINGE (2), PCS MISSING (10). LOT 9213635 - PC PRE-CUT (30), PCS STAINS ON PACKAGING (34), PCS STAINS ON SYRINGE (1).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1182337 | BD POSIFLUSH¿ XS PRE-FILLED FLUSH SYRINGE NACL 0.9% | SALINE, VASCULAR ACCESS FLUSH | NGT | BECTON, DICKINSON AND CO. | SEE H.10 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |