FDA Adverse Event Malfunction Summary report: N

SQUARE SCREWDRIVER

MDR report key: 9378968 · Received November 26, 2019

Report

Report Number
3005985723-2019-00856
Event Type
Malfunction
Date Received
November 26, 2019
Date of Event
November 5, 2019
Report Date
February 3, 2020
Manufacturer
MAKO SURGICAL CORP.
Product Code
OLO
UDI-DI
00848486002763
PMA / PMN Number
K112507
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

REPORTED EVENT: -CUSTOMER REPORTED THAT THE SILVER END OF SQUARE DRIVER CAME LOOSE FROM THE BLUE HANDLE WHILE TAKING APART ARRAYS/ARRAY CLAMPS. DEVICE EVALUATION AND RESULTS: DEVICE EVALUATION WAS PERFORMED AND THE SQUARE SCREWDRIVER EVENT WAS CONFIRMED. DEVICE HISTORY REVIEW: REVIEW OF THE DEVICE HISTORY RECORDS INDICATE 379 DEVICES WERE MANUFACTURED AND INSPECTED ON 07-02-2018 WITH NO REPORTED DISCREPANCIES. COMPLAINT HISTORY REVIEW: BASED ON THE DEVICE IDENTIFICATION, THE CATSWEB AND TRACKWISE COMPLAINT DATABASES WERE REVIEWED FROM 2011 TO PRESENT FOR SIMILAR REPORTED EVENTS REGARDING THE SQUARE SCREWDRIVER. THERE HAVE BEEN 50 OTHER EVENTS FOR THE REFERENCED LOT NUMBER. PRS # 1914925, 1919787, 1935822, 1943690, 1947954, 1953514, 1967031, 1970724, 1987510, 1978669, 1995528, 2001610, 2017816, 2018871, 2019846, 2022521, 2028928, 2030818, 2031968, 2036455, 2045046, 2045740, 2051183, 2073709, 2082053, 2103869, 2114279, 2114688, 2117018, 2122244, 2125456, 2127852, 2139397, 2149504, 215096, 2154940, 21666790, 2188146, 2188148, 2192078, 2204673, 2208809, 2209518, 2210264 , 2210272, 2211307, 2213526, 2221304, 2221688 AND 2228967 - WERE FOUND TO BE FROM THE SAME LOT AS THE PART IN THIS COMPLAINT. THE PARTS FROM THESE PRS DISPLAYED THE SAME FAILURE. CONCLUSIONS: THE SQUARE SCREWDRIVER SHOWED THE MASTERBOND THAT HOLDS THE SHAFT AND HANDLE CRACKED. CORRECTIVE ACTION/PREVENTIVE ACTION: A REVIEW OF STRYKER¿S NC/CAPA DATABASE INDICATED THAT CAPA 1694870 WAS INITIATED FOR THIS PART NUMBER.

Description of Event or Problem · 0

SILVER END OF SQUARE DRIVER CAME LOOSE FROM THE BLUE HANDLE WHILE TAKING APART ARRAYS/ARRAY CLAMPS. CASE TYPE: PKA (ANSPACH). SURGICAL DELAY: > 30 MINUTES.

Additional Manufacturer Narrative · 1

AS PART OF THE NORMAL COMPLAINT FOLLOW-UP, AN EVALUATION OF THE EVENT HAS BEEN INITIATED BY MAKO SURGICAL. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE.

Description of Event or Problem · 1

SILVER END OF SQUARE DRIVER CAME LOOSE FROM THE BLUE HANDLE WHILE TAKING APART ARRAYS/ARRAY CLAMPS. CASE TYPE: PKA (ANSPACH). SURGICAL DELAY: > 30 MINUTES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1172880 SQUARE SCREWDRIVER STEREOTAXIC DEVICE, ROBOTICS OLO MAKO SURGICAL CORP. 111160 06110518 00848486002763

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization