COMET
Report
- Report Number
- 2134265-2019-14052
- Event Type
- Malfunction
- Date Received
- November 26, 2019
- Date of Event
- September 21, 2019
- Report Date
- November 26, 2019
- Manufacturer
- BOSTON SCIENTIFIC CORPORATION
- Product Code
- DXO
- UDI-DI
- 08714729904397
- PMA / PMN Number
- K151610
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO
- Reporter Occupation
- PHYSICIAN
Narratives
DEVICE RETURNED TO MANUFACTURER: THE COMET WAS RETURNED AND ANALYSIS WAS COMPLETED. THE TIP, DEVICE SHAFT, SENSOR PORT AND THE COEFFICIENT VALUES WERE EXAMINED FOR DAMAGE OR ANY IRREGULARITIES. THE SHAFT SHOWED TWO KINKS. THE FIRST KINK WAS LOCATED 38CM FROM THE TIP. THE SECOND KINK WAS LOCATED 177CM FROM THE TIP. THERE WAS TIP DAMAGE NOTICED. THERE WAS PEELED COATING AT THE 177CM LOCATION. THE WIRE WAS CONNECTED TO THE RETURNED OCC HANDLE. THE OCC HANDLE WAS CONNECTED TO THE FFR LINK FOR SIGNAL VERIFICATION. THE SIGNAL WAS NOT PRESENT AS DESIGNED. THE OCC HANDLE CAP WAS LOOSENED TO REMOVE THE WIRE. THERE WAS NO ISSUE WITH REMOVING THE WIRE. A KNOWN GOOD OCC HANDLE WAS USED TO VERIFY CABLE OPERATION WAS CORRECT WITH THE RETURNED OCC. THE TEST OCC CABLE AND THE CONNECTED RETURNED WIRE FUNCTIONED AS DESIGNED. THE RETURNED CABLE WAS DISSECTED TO CHECK FOR DAMAGE. IT WAS NOTICED THAT THE FEP DISK WAS DAMAGED WHERE THE WIRE SEATS AGAINST THE DISK. INSPECTION OF THE REMAINDER OF THE DEVICE, APART FROM THE OBSERVED DAMAGE, REVEALED NO OTHER DAMAGE OR IRREGULARITIES.
REPORTABLE BASED ON ANALYSIS COMPLETED 14 NOV 2019. IT WAS REPORTED THAT THE COMET PRESSURE HAD BEEN UNABLE TO BE RECOGNIZED BY THE FFR LINK. ANOTHER COMET WAS USED TO COMPLETE THE PROCEDURE WITHOUT ISSUE. HOWEVER, THE RETURNED DEVICE ANALYSIS REVEALED SOME SLIGHT PEELING OF THE DEVICE COATING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1172036 | COMET | TRANSDUCER, PRESSURE, CATHETER TIP | DXO | BOSTON SCIENTIFIC CORPORATION | 8900 | 0023099681 | 08714729904397 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |