FDA Adverse Event Malfunction Summary report: N

HANDPIECE MICS

MDR report key: 9378679 · Received November 26, 2019

Report

Report Number
3005985723-2019-00854
Event Type
Malfunction
Date Received
November 26, 2019
Date of Event
November 4, 2019
Report Date
February 27, 2020
Manufacturer
MAKO SURGICAL CORP.
Product Code
OLO
UDI-DI
00848486030193
PMA / PMN Number
K142530
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

TRIGGER ON MICS IS STICKY AND GETS STUCK MAKING IT VERY DIFFICULT TO ENGAGE HAPTICS AND COMPLETE CUTS. CASE TYPE: TKA. SURGICAL DELAY: YES > 30 MINUTES. PRODUCT INSPECTION AS PER SERVICE MAXWO-01688588 & CASE NUMBER (B)(4) SN # (B)(6). FACTORY SERVICE TECHNICIAN: (B)(6). 1. SUCCESSFULLY PERFORMED MICS CABLE REPLACEMENT PER DO6917. MICS CAN BE RETURNED TO STOCK AS A 209063 LOT #4201501 -R. THE ALLEGED FAILURE WAS CONFIRMED DEVICE HISTORY REVIEW: DEVICE HISTORY RECORDS INDICATE (B)(4) DEVICES WERE MANUFACTURED UNDER L/N 42021216 AND ALL (B)(4) DEVICES WERE REJECTED ON (B)(6) 2017. A REVIEW OF QT17-01-0035 REVEALED THAT THEY WERE REJECTED DUE TO DOCUMENTATION ISSUE. UNITS WERE ACCEPTED INTO FINAL STOCK AFTER DOCUMENTATION WAS PROVIDED. THE NON CONFORMANCE IS NOT RELATED TO THE FAILURE ALLEGED IN THIS COMPLAINT. COMPLAINT HISTORY REVIEW: A REVIEW OF COMPLAINTS IN CATSWEB AND TRACKWISE RELATED TO P/N 209063, LOT NUMBER 42021216 SHOWS 01 ADDITIONAL COMPLAINT RELATED TO THE FAILURE IN THIS INVESTIGATION. CONCLUSION: THE ALLEGED FAILURE MODE WAS CONFIRMED. NO ADDITIONAL INVESTIGATION OR SPECIFIC ACTIONS ARE REQUIRED.

Description of Event or Problem · 0

TRIGGER ON MICS IS STICKY AND GETS STUCK MAKING IT VERY DIFFICULT TO ENGAGE HAPTICS AND COMPLETE CUTS. CASE TYPE: TKA. SURGICAL DELAY: YES > 30 MINUTES.

Additional Manufacturer Narrative · 1

AS PART OF THE NORMAL COMPLAINT FOLLOW-UP, AN EVALUATION OF THE EVENT HAS BEEN INITIATED BY MAKO SURGICAL. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE.

Description of Event or Problem · 1

TRIGGER ON MICS IS STICKY AND GETS STUCK MAKING IT VERY DIFFICULT TO ENGAGE HAPTICS AND COMPLETE CUTS. CASE TYPE: TKA. SURGICAL DELAY: YES > 30 MINUTES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1169683 HANDPIECE MICS STEREOTAXIC DEVICE, ROBOTICS OLO MAKO SURGICAL CORP. 209063 4201501 00848486030193

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization