FDA Adverse Event Malfunction Summary report: N

AIR PEN DRIVE

MDR report key: 9378591 · Received November 26, 2019

Report

Report Number
8030965-2019-70520
Event Type
Malfunction
Date Received
November 26, 2019
Date of Event
January 1, 2019
Report Date
November 12, 2019
Manufacturer
DEPUY SYNTHES PRODUCTS LLC
Product Code
HBB
UDI-DI
07611819830899
PMA / PMN Number
K093361
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THIS MEDWATCH, A SUPPLEMENTAL MEDWATCH REPORT WILL BE FILED AS APPROPRIATE. THE ACTUAL DEVICE WAS RETURNED FOR EVALUATION. DURING REPAIR, AN EVALUATION WAS PERFORMED AND IT WAS DETERMINED THAT THE DEVICE HAD INSUFFICIENT/LOW POWER. IT WAS FURTHER DETERMINED THAT THE DEVICE FAILED PRETEST FOR CHECK POWER WITH POWER TEST BENCH, CHECK SPEED WITH TACHOMETER, CHECK EXTERNAL LEAK TIGHTNESS, CHECK INTERNAL LEAK TIGHTNESS, AND MARKING AND LABELING. THEREFORE, THE REPORTED CONDITION THAT THE DEVICE HAD INSUFFICIENT POWER CAUSING THE DEVICE TO RUN SLOW AND WITH LOWER ENERGY AND LOW ROTATIONAL SPEED WAS CONFIRMED. THE ASSIGNABLE ROOT CAUSE OF THIS CONDITION WAS DETERMINED TO BE TRACED TO COMPONENT FAILURE DUE TO NORMAL WEAR.

Additional Manufacturer Narrative · 0

H10: DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. H10 ADDITIONAL INFORMATION: G1, H4: THE MANUFACTURING SITE NAME WAS DOCUMENTED AS UNKNOWN IN THE INITIAL MEDWATCH REPORT. THIS HAS BEEN UPDATED TO SYNTHES PRODUKTIONS GMBH (WALDENBURG). PLEASE NOTE THAT THE CONTACT OFFICE NAME/ADDRESS HAVE BEEN UPDATED ACCORDINGLY TO REFLECT THE CORRECT MANUFACTURING FACILITY. THE DATE OF MANUFACTURE (DOM) WAS DOCUMENTED AS UNKNOWN IN THE INITIAL MEDWATCH REPORT. THE DOM HAS BEEN UPDATED TO SEPTEMBER 17, 2007. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

Additional Manufacturer Narrative · 1

DEVICE MANUFACTURE DATE: THE DEVICE MANUFACTURE DATE IS UNAVAILABLE. THE MANUFACTURING LOCATION WAS UNKNOWN. REPORTER'S PHONE NUMBER WAS NOT PROVIDED. AS OF THIS DATE, THE DEVICE HAS NOT BEEN RETURNED FOR EVALUATION; THEREFORE, THE REPORTED CONDITION CANNOT BE CONFIRMED AND/OR DUPLICATED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. UDI: (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED FROM (B)(6) THAT THE AIR PEN DRIVE DEVICE HAD INSUFFICIENT POWER CAUSING THE DEVICE TO RUN SLOW AND WITH LOWER ENERGY AND LOW ROTATIONAL SPEED. IT WAS NOT REPORTED IF THE EVENT OCCURRED DURING A SURGICAL PROCEDURE. IT WAS NOT REPORTED IF THERE WAS A DELAY TO A PLANNED PROCEDURE. IT WAS NOT REPORTED IF THERE WAS A SPARE DEVICE AVAILABLE FOR USE. THERE WAS NO PATIENT INVOLVEMENT REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS NOT REPORTED, HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN 2019. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1169087 AIR PEN DRIVE MOTOR, DRILL, PNEUMATIC HBB DEPUY SYNTHES PRODUCTS LLC 07611819830899

Patients

Seq Age Sex Outcome Treatment
1