FDA Adverse Event Injury Summary report: N

NOVOSYN VIOLET 0 (3.5) 70CM FR26 (M)

MDR report key: 9377307 · Received November 26, 2019

Report

Report Number
3003639970-2019-00848
Event Type
Injury
Date Received
November 26, 2019
Date of Event
October 23, 2019
Report Date
January 3, 2020
Manufacturer
B.BRAUN SURGICAL SA
Product Code
GAM
PMA / PMN Number
SEE H10
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION: SAMPLES RECEIVED: 1 OPEN UNIT. ANALYSIS AND RESULTS: THERE ARE NO PREVIOUS COMPLAINTS OF THIS CODE-BATCH. WE MANUFACTURED AND DISTRIBUTED IN THE MARKET (B)(4) UNITS OF THIS CODE-BATCH. THERE ARE NO UNITS IN OUR STOCK. WE HAVE RECEIVED ONE OPEN AND UNUSED SAMPLE WITH THE NEEDLE DETACHED FROM THE THREAD (THREAD IS STILL WOUND ON THE PACK). HOWEVER, WITHOUT CLOSED SAMPLE WE CANNOT CARRY OUT AN ANALYSIS IN ORDER TO TAKE A DECISION. REVIEWED THE BATCH MANUFACTURING RECORD, THIS PRODUCT HAD A NORMAL PROCESS AND THE RESULTS DURING THE PROCESS FULFIL USP/EP AND B. BRAUN SURGICAL REQUIREMENTS. TAKING INTO ACCOUNT THAT NO OTHER CUSTOMER COMPLAINTS HAVE BEEN RECEIVED FOR THIS CODE-BATCH, WE CONSIDER THAT THIS IS AN ISOLATED UNIT, BUT THE WHOLE BATCH IS CORRECT. FINAL CONCLUSION: IN SPITE OF RECEIVING A DEFECTIVE SAMPLE, WITHOUT CLOSED SAMPLES A SUITABLE ANALYSIS CANNOT BE PERFORMED. NEVERTHELESS, WE TAKE NOTE OF THIS INCIDENCE AND IF ANY CLOSED SAMPLE IS RECEIVED IN THE FUTURE, WE WILL RE-OPEN THE CASE AND ANALYZE IT. ACTIONS ON PRODUCT DISTRIBUTED OF THIS REFERENCE/BATCH: BASED ON THE CONCLUSION DERIVED FROM INVESTIGATION, IT IS NOT REQUIRED TO MAKE ACTIONS IN DISTRIBUTED PRODUCT. NO CORRECTIVE/PREVENTIVE ACTIONS NEEDED.

Additional Manufacturer Narrative · 1

PMA/510K: REPORTED DEVICE NOT MARKETED IN THE U.S., HOWEVER, SIMILAR DEVICES OR DEVICES THAT SHARE COMPONENTS, RAW MATERIALS, PROCESS METHODS OR OTHER TECHNOLOGICAL CHARACTERISTICS ARE REGISTERED WITHIN THE U.S. K122734. WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE A FOLLOW UP REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED NEEDLE DETACHES FROM THE THREAD. THE REPORTER INDICATED THAT THE NEEDLE DETACHED FROM THE THREAD. ADDITIONAL INFORMATION HAS BEEN REQUESTED, HOWEVER, NOT YET RECEIVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1173372 NOVOSYN VIOLET 0 (3.5) 70CM FR26 (M) SYNTHETIC ABSORBABLE BRAIDED S GAM B.BRAUN SURGICAL SA C0068416 119282

Patients

Seq Age Sex Outcome Treatment
1