NOVOSYN VIOLET 0 (3.5) 70CM FR26 (M)
Report
- Report Number
- 3003639970-2019-00848
- Event Type
- Injury
- Date Received
- November 26, 2019
- Date of Event
- October 23, 2019
- Report Date
- January 3, 2020
- Manufacturer
- B.BRAUN SURGICAL SA
- Product Code
- GAM
- PMA / PMN Number
- SEE H10
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
INVESTIGATION: SAMPLES RECEIVED: 1 OPEN UNIT. ANALYSIS AND RESULTS: THERE ARE NO PREVIOUS COMPLAINTS OF THIS CODE-BATCH. WE MANUFACTURED AND DISTRIBUTED IN THE MARKET (B)(4) UNITS OF THIS CODE-BATCH. THERE ARE NO UNITS IN OUR STOCK. WE HAVE RECEIVED ONE OPEN AND UNUSED SAMPLE WITH THE NEEDLE DETACHED FROM THE THREAD (THREAD IS STILL WOUND ON THE PACK). HOWEVER, WITHOUT CLOSED SAMPLE WE CANNOT CARRY OUT AN ANALYSIS IN ORDER TO TAKE A DECISION. REVIEWED THE BATCH MANUFACTURING RECORD, THIS PRODUCT HAD A NORMAL PROCESS AND THE RESULTS DURING THE PROCESS FULFIL USP/EP AND B. BRAUN SURGICAL REQUIREMENTS. TAKING INTO ACCOUNT THAT NO OTHER CUSTOMER COMPLAINTS HAVE BEEN RECEIVED FOR THIS CODE-BATCH, WE CONSIDER THAT THIS IS AN ISOLATED UNIT, BUT THE WHOLE BATCH IS CORRECT. FINAL CONCLUSION: IN SPITE OF RECEIVING A DEFECTIVE SAMPLE, WITHOUT CLOSED SAMPLES A SUITABLE ANALYSIS CANNOT BE PERFORMED. NEVERTHELESS, WE TAKE NOTE OF THIS INCIDENCE AND IF ANY CLOSED SAMPLE IS RECEIVED IN THE FUTURE, WE WILL RE-OPEN THE CASE AND ANALYZE IT. ACTIONS ON PRODUCT DISTRIBUTED OF THIS REFERENCE/BATCH: BASED ON THE CONCLUSION DERIVED FROM INVESTIGATION, IT IS NOT REQUIRED TO MAKE ACTIONS IN DISTRIBUTED PRODUCT. NO CORRECTIVE/PREVENTIVE ACTIONS NEEDED.
PMA/510K: REPORTED DEVICE NOT MARKETED IN THE U.S., HOWEVER, SIMILAR DEVICES OR DEVICES THAT SHARE COMPONENTS, RAW MATERIALS, PROCESS METHODS OR OTHER TECHNOLOGICAL CHARACTERISTICS ARE REGISTERED WITHIN THE U.S. K122734. WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE A FOLLOW UP REPORT WILL BE SUBMITTED.
IT WAS REPORTED NEEDLE DETACHES FROM THE THREAD. THE REPORTER INDICATED THAT THE NEEDLE DETACHED FROM THE THREAD. ADDITIONAL INFORMATION HAS BEEN REQUESTED, HOWEVER, NOT YET RECEIVED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1173372 | NOVOSYN VIOLET 0 (3.5) 70CM FR26 (M) | SYNTHETIC ABSORBABLE BRAIDED S | GAM | B.BRAUN SURGICAL SA | C0068416 | 119282 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |