FDA Adverse Event Malfunction Summary report: N

JACKSON-PRATT CHANNEL DRAIN

MDR report key: 9375035 · Received November 26, 2019

Report

Report Number
8030107-2019-00022
Event Type
Malfunction
Date Received
November 26, 2019
Date of Event
July 19, 2019
Report Date
November 26, 2019
Manufacturer
DEGANIA SILICONE LTD
Product Code
GCB
PMA / PMN Number
EXEMPT
Removal / Correction Number
8030107-06/28/2019-001-R
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
IS
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE LOT NUMBER OF THE BROKEN DEVICE IS UNKNOWN; THE DEVICE PROBABLY BELONGS TO ONE OF THE LOTS AFFECTED BY RECALL. IN 2019 WE OBSERVE HAVE SHARP INCREASE OF REPORTS ON BREAKAGE IN THESE DRAINS DURING USE, IN CONNECTION AREA. THE BREAKAGE WAS ASSOCIATED WITH THE NEW TOOL USED FOR THE PRODUCTION OF CONNECTOR WHICH CONNECTS CLEAR TUBE AND WHITE DRAINING PORTION IN THESE DRAINS. FOLLOWING THIS INCREASED RATE, ON 06/28/2019, DEGANIA INITIATED RECALL #8030107-06/28/2019-001-R TO REMOVE ALL DRAINS IN WHICH NEW DESIGN OF CONNECTORS WAS USED.

Description of Event or Problem · 1

AFTER WHIPPLE PROCEDURE, THE PATIENT CAME FOR AN OFFICE VISIT FOR RIGHT UPPER QUADRANT PAIN AND DECREASED DRAINAGE FROM THE JP TUBE. AN ABDOMINAL CT NOTED A SEPARATION OF THE CLEAR PORTION OF THE JP DRAIN FROM THE WHITE PORTION. PATIENT UNDERWENT A DIAGNOSTIC LAPAROSCOPY FOR REMOVAL OF RETAINED PORTION OF THE DRAIN, AND REPLACEMENT OF SECOND DRAIN. THE WHITE PORTION OF THE DRAIN WAS TUNNELING OUT BETWEEN LOOPS OF BOWEL AND THERE WAS CLOUDY PANCREATIC TYPE FLUID COMING OUT FROM THIS TRACT ONCE THE FLUTED DRAIN WAS REMOVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1170314 JACKSON-PRATT CHANNEL DRAIN SILICONE FLUTED 15FR DRAIN GCB DEGANIA SILICONE LTD JP-2189 UNK

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention