FDA Adverse Event Malfunction Summary report: N

NAV THORACIC PROBE TIP

MDR report key: 9371937 · Received November 25, 2019

Report

Report Number
1723170-2019-05772
Event Type
Malfunction
Date Received
November 25, 2019
Date of Event
November 6, 2019
Report Date
February 5, 2020
Manufacturer
MEDTRONIC NAVIGATION, INC
Product Code
OLO
UDI-DI
00763000054205
PMA / PMN Number
K124004
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

PATIENT INFORMATION WAS UNAVAILABLE FROM THE SITE. NO PARTS HAVE BEEN RETURNED TO THE MANUFACTURER FOR ANALYSIS. INITIAL REPORTER UNKNOWN AT THE TIME OF THIS REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

ADDITIONAL INFORMATION WAS RECEIVED AND HAS BEEN UPDATED. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

THE THORACIC PROBE 9734680 NAVLOCK (LOT# 190516) WAS RETURNED FOR ANALYSIS. ANALYSIS FOUND THAT THE RETURNED PROBE HAD SEVERAL IMPACT MARKS AT THE BACK END CAUSING THE REPORTED FIT ISSUES WITH THE MATING TRACKER. THE HARDWARE INVESTIGATION FOUND THAT THE REPORTED EVENT WAS RELATED TO A HARDWARE ISSUE. THIS ISSUE WAS DOCUMENTED IN A MEDTRONIC NAVIGATION HARDWARE ANOMALY TRACKING DATABASE. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

MEDTRONIC RECEIVED INFORMATION REGARDING NAVIGATION DEVICES USED FOR SACROILIAC AND THORACOLUMBAR PROCEDURES. IT WAS REPORTED THAT THREE PROBES WERE ALL DAMAGED IN THE SAME SPOT. IT WAS NOTED THAT THE CLINICAL DETAILS OF THE CASES WERE UNKNOWN BY THE SITE AND THAT THE SITE USED A MALLET ON THE INSTRUMENTS. THERE WAS A LESS THAN HOUR SURGICAL DELAY AND NO REPORTED IMPACT ON PATIENT OUTCOME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1165035 NAV THORACIC PROBE TIP ORTHOPEDIC STEREOTAXIC INSTRUMENT OLO MEDTRONIC NAVIGATION, INC 9734680 190516 00763000054205

Patients

Seq Age Sex Outcome Treatment
1