FDA Adverse Event Malfunction Summary report: N

SMITHS MEDICAL BIVONA TUBES NEO/PED HYPERFLEX

MDR report key: 9371228 · Received November 25, 2019

Report

Report Number
3012307300-2019-06798
Event Type
Malfunction
Date Received
November 25, 2019
Date of Event
November 18, 2019
Report Date
May 7, 2020
Manufacturer
SMITHS MEDICAL ASD, INC.
Product Code
JOH
UDI-DI
15021312013702
PMA / PMN Number
K081440
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
501

Narratives

Additional Manufacturer Narrative · 0

RETURNED DEVICE WAS RECEIVED IN GOOD PHYSICAL CONDITION. DURING THE EVALUATION OF THE DEVICE, IT WAS FOUND THAT THE DEVICE COULD MOVE FREELY AND THE LOCKING MECHANISM WORKED JUST FINE. THERE IS SPECULATION THAT THE DEFECT WAS CORRECTLY COMMUNICATED OR THE CORRECT DEVICE WAS RETURNED. ADDITIONALLY SINCE THE NECK FLANGE ROTATES AROUND THE SHAFT, REPOSITIONING THE NECK FLANGE SO THE FLANGES ARE NOT UPSIDE DOWN WOULD HAVE CORRECTED THE CONCERN AND TYPICALLY NOT WARRANTED A COMPLAINT. THE DEVICE HISTORY RECORD WAS REVIEWED AND SHOWED THAT THIS DEVICE MET ALL MANUFACTURING SPECIFICATION FOR PRODUCT RELEASED FOR DISTRIBUTION. NO ISSUES WERE IDENTIFIED THAT WOULD HAVE IMPACTED THIS EVENT.

Additional Manufacturer Narrative · 0

DEVICE EVALUATION IN PROGRESS.

Description of Event or Problem · 0

INFORMATION WAS RECEIVED THAT A SMITHS MEDICAL BIVONA HYPERFLEX TRACHEOSTOMY TUBE HAD A NECK PLATE ON UPSIDE DOWN. AS A RESULT, THE PATIENT WAS NOT PROPERLY VENTILATED, AND A NEW TRACHEOSTOMY WAS INSERTED. IT WAS FURTHER REPORTED THAT THE USER NOTICED UNSPECIFIED PROBLEMS WITH TWO ADDITIONAL BIVONA HYPERFLEX TRACHEOSTOMY TUBES. THESE UNSPECIFIED PRODUCT ISSUES WERE IDENTIFIED PRIOR TO PATIENT USE.

Description of Event or Problem · 1

INFORMATION WAS RECEIVED THAT A SMITHS MEDICAL BIVONA HYPERFLEX TRACHEOSTOMY TUBE HAD A NECK PLATE ON UPSIDE DOWN. AS A RESULT, THE PATIENT WAS NOT PROPERLY VENTILATED, AND A NEW TRACHEOSTOMY WAS INSERTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1164321 SMITHS MEDICAL BIVONA TUBES NEO/PED HYPERFLEX TUBE TRACHEOSTOMY AND TUBE CUFF JOH SMITHS MEDICAL ASD, INC. 60HA35 38001467 15021312013702

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention