FDA Adverse Event
Malfunction
Summary report: N
HORIZON MEDICAL IMAGING
MDR report key: 937082
·
Received October 31, 2007
Report
- Report Number
- 8022257-2007-00003
- Event Type
- Malfunction
- Date Received
- October 31, 2007
- Date of Event
- June 13, 2007
- Report Date
- October 31, 2007
- Manufacturer
- MCKESSON MEDICAL IMAGING COMPANY
- Product Code
- LLZ
- PMA / PMN Number
- K043146
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
MCKESSON MEDICAL IMAGING COMPANY DETERMINED THAT NON-STANDARD CONFIGURATION SETTINGS ON THE LEGACY PACS AND HORIZON MEDICAL IMAGING (HMI) AT SPECTRUM HEALTH CONTRIBUTED TO THE REPORTED PROBLEM. THE PROBLEM WAS RESOLVED BY RETURNING THE PARTICULAR CONFIGURATION VARIABLES TO THEIR DEFAULT SETTINGS. IT WAS DETERMINED THAT THE HMI SYSTEM AT SPECTRUM HEALTH WAS THE ONLY CUSTOMER WHICH HAD CONFIGURATION SETTINGS THAT WOULD GIVE RISE TO THE REPORTED PROBLEM.
Description of Event or Problem · 1
SOME STUDIES STORED ONLINE IN MEDICAL IMAGING FROM DICOM PUSH IMPORT FROM A CUSTOMER'S LEGACY PACS, WHICH HAD NON-STANDARD CONFIGURATION SETTINGS, MAY BE ASSOCIATED WITH AN INCORRECT PATIENT AND MISIDENTIFIED AS BELONGING TO THAT PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HORIZON MEDICAL IMAGING | PICTURE ARCHIVING AND COMMUNICATION SYS. | LLZ | MCKESSON MEDICAL IMAGING COMPANY | HMI 11.0 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | YR |