FDA Adverse Event Injury Summary report: N

MAESTRO RAD W/BRG 7X15 7.5 RT

MDR report key: 9370562 · Received November 25, 2019

Report

Report Number
0001825034-2019-05314
Event Type
Injury
Date Received
November 25, 2019
Report Date
November 25, 2019
Manufacturer
ZIMMER BIOMET, INC.
Product Code
JWJ
UDI-DI
00880304439092
PMA / PMN Number
K042032
Removal / Correction Number
N/A
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). CONCOMITANT MEDICAL PRODUCTS: CATALOG #: 180180, MAESTRO RADIAL STEM 4.5X40MM, LOT # 181430, CATALOG #: 180351, VARIABLE LOCK SCREW 4.75X20MM, LOT # 748570, CATALOG #: 180363, MAESTRO TC CARPALHD 7X15MM STD, LOT # 527190, CATALOG #: 180352, VARIABLE LOCK SCREW 4.75X25MM, LOT # 551330, CATALOG #: 180363, MAESTRO TC CARPALHD 7X15MM STD, LOT # 527190, CATALOG #: 180321, MAESTRO TC CAPITATESTEM 6X15MM, LOT # 238690N THE CUSTOMER HAS NOT INDICATED WHETHER THE PRODUCT WILL BE RETURNED TO ZIMMER BIOMET FOR INVESTIGATION OR NOT. ONCE THIS INFORMATION IS OBTAINED A FOLLOW-UP MDR WILL BE SUBMITTED. REPORTED EVENT WAS CONSIDERED CONFIRMED DUE TO THE PROVED X-RAYS THAT SHOWED LOOSENING OF THE RADIAL COMPONENT. DEVICE HISTORY RECORD WAS REVIEWED AND NO DISCREPANCIES RELEVANT TO THE REPORTED EVENT WERE FOUND. ROOT CAUSE WAS CONSIDERED TO BE USER ERROR AS IT WAS KNOWN THAT THE PATIENT WAS A NURSE THAT LIFTED PATIENTS. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS. MULTIPLE MDR REPORTS WERE FILED FOR THIS EVENT, PLEASE SEE ASSOCIATED REPORTS: 0001825034-2019-04291.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT IS SCHEDULED TO BE GETTING A REVISION DUE TO THE RADIAL STEM BEING LOOSE IN THE CANAL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1165770 MAESTRO RAD W/BRG 7X15 7.5 RT PROSTHESIS SHOULDER JWJ ZIMMER BIOMET, INC. N/A 922440 00880304439092

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| R