FDA Adverse Event Malfunction Summary report: N

BD ECLIPSE¿ 3 ML BD LUER-LOK¿ SYRINGE

MDR report key: 9369906 · Received November 25, 2019

Report

Report Number
8041187-2019-00945
Event Type
Malfunction
Date Received
November 25, 2019
Date of Event
November 6, 2019
Report Date
December 11, 2019
Manufacturer
BECTON DICKINSON MEDICAL (SINGAPORE)
Product Code
FMI
UDI-DI
30382903057833
PMA / PMN Number
K980987
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION: ONE SAMPLE WAS RECEIVED BY OUR QUALITY TEAM FOR EVALUATION. UPON VISUAL INSPECTION THERE WAS NO HUB BREAKAGE OR SAFETY SHIELD FALL OFF; THEREFORE, THE INCIDENT CAN NOT BE VERIFIED. A DEVICE HISTORY RECORD REVIEW FOUND NO NON- CONFORMANCES ASSOCIATED WITH THIS ISSUE DURING PRODUCTION OF THIS BATCH. BASED ON THE QUALITY TEAM'S INVESTIGATION THE PROBABLE ROOT CAUSE COULD NOT BE DETERMINED. H3 OTHER TEXT : SEE H.10.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD ECLIPSE¿ 3 ML BD LUER-LOK¿ SYRINGE SAFETY MECHANISM DETACHED. THIS WAS DISCOVERED DURING USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL NO: 305783, BATCH NO: 9081645. IT WAS REPORTED THAT WHEN NURSE ATTEMPTED TO ACTIVATE SAFETY MECHANISM, THE ARM DISENGAGED FROM THE HINGES AND FELL OFF. NURSE ATTEMPTED TO ACTIVATE THE SAFETY MECHANISM AND THE ARM DISENGAGED FROM THE HINGES AND FELL OFF.

Additional Manufacturer Narrative · 1

A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT BD ECLIPSE¿ 3 ML BD LUER-LOK¿ SYRINGE SAFETY MECHANISM DETACHED. THIS WAS DISCOVERED DURING USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL NO: 305783, BATCH NO: 9081645. IT WAS REPORTED THAT WHEN NURSE ATTEMPTED TO ACTIVATE SAFETY MECHANISM, THE ARM DISENGAGED FROM THE HINGES AND FELL OFF. NURSE ATTEMPTED TO ACTIVATE THE SAFETY MECHANISM AND THE ARM DISENGAGED FROM THE HINGES AND FELL OFF.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1168271 BD ECLIPSE¿ 3 ML BD LUER-LOK¿ SYRINGE HYPODERMIC SINGLE LUMEN NEEDLE FMI BECTON DICKINSON MEDICAL (SINGAPORE) 9081645 30382903057833

Patients

Seq Age Sex Outcome Treatment
1 Other