FDA Adverse Event Injury Summary report: N

PREMILENE 3/0 (2) 90CM 2XHR26 CV

MDR report key: 9369437 · Received November 25, 2019

Report

Report Number
3003639970-2019-00813
Event Type
Injury
Date Received
November 25, 2019
Report Date
January 3, 2020
Manufacturer
B.BRAUN SURGICAL SA
Product Code
GAW
PMA / PMN Number
SEE H10
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION: SAMPLES RECEIVED: 15 UNOPENED RACE-PACKS. ANALYSIS AND RESULTS: THERE ARE NO PREVIOUS COMPLAINTS OF THIS CODE BATCH OF WHICH WE MANUFACTURED AND DISTRIBUTED IN THE MARKET (B)(4) UNITS. THERE ARE NO UNITS IN STOCK IN B. BRAUN SURGICAL'S WAREHOUSE. WE HAVE RECEIVED 15 CLOSED SAMPLES. WE HAVE TESTED THE KNOT PULL TENSILE STRENGTH OF THE CLOSED SAMPLES RECEIVED AND THE RESULTS FULFILL THE REQUIREMENTS OF THE EUROPEAN PHARMACOPOEIA (EP): 1.72 KGF IN AVERAGE AND 1.62 KGF IN MINIMUM (EP REQUIREMENTS: 0.92 KGF IN AVERAGE AND 0.31 KGF IN MINIMUM). REVIEWED THE BATCH MANUFACTURING RECORD, THIS PRODUCT HAD A NORMAL PROCESS AND THE RESULTS DURING THE PROCESS FULFILL USP/EP AND B. BRAUN SURGICAL REQUIREMENTS. FINAL CONCLUSION: ALTHOUGH THE RESULTS OF THE SAMPLES RECEIVED FULFIL THE SPECIFICATIONS OF EUROPEAN PHARMACOPOEIA/ B. BRAUN SURGICAL SPECIFICATIONS, WE TAKE NOTE OF THIS INCIDENCE IN ORDER TO ASSESS IF NEW OR ADDITIONAL ACTIONS ARE NEEDED. THE CASE IS CONSIDERED NOT CONFIRMED BY EVIDENCE OF THE SAMPLES RECEIVED. ACTIONS ON PRODUCT DISTRIBUTED OF THIS REFERENCE/BATCH: BASED ON THE CONCLUSION DERIVED FROM INVESTIGATION, IT IS NOT REQUIRED TO MAKE ACTIONS IN DISTRIBUTED PRODUCT. NO CORRECTIVE/PREVENTIVE ACTIONS NEEDED.

Additional Manufacturer Narrative · 1

PMA/510K: REPORTED DEVICE NOT MARKETED IN THE U.S., HOWEVER, SIMILAR DEVICES OR DEVICES THAT SHARE COMPONENTS, RAW MATERIALS, PROCESS METHODS OR OTHER TECHNOLOGICAL CHARACTERISTICS ARE REGISTERED WITHIN THE U.S. K980703. IF ADDITIONAL INFORMATION BECOMES AVAILABLE A FOLLOW UP REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THE SUTURE BREAKS. THE REPORTER INDICATED THAT WHEN THE SUTURE IS SUTURED, THE SUTURE BREAKS. THE REPORTER STATED THIS EVENT OCCURRED PRIOR TO USE ON A PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1162461 PREMILENE 3/0 (2) 90CM 2XHR26 CV CARDIAC SUTURE GAW B.BRAUN SURGICAL SA C2090908 117395

Patients

Seq Age Sex Outcome Treatment
1