FDA Adverse Event Malfunction Summary report: N

BD 1 ML SYRINGE LUER SLIP TIP CONVENIENCE PAK

MDR report key: 9365489 · Received November 22, 2019

Report

Report Number
9610847-2019-00697
Event Type
Malfunction
Date Received
November 22, 2019
Date of Event
November 5, 2019
Report Date
January 27, 2020
Manufacturer
BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V.
Product Code
FMF
UDI-DI
30382903097013
PMA / PMN Number
K980580
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
PHARMACIST

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION: ONE CONVENIENCE TRAY OF TWENTY-FIVE SYRINGES WAS PROVIDED FOR INVESTIGATION BY OUR QUALITY ENGINEER TEAM. THROUGH EXAMINATION OF THE RETURNED SAMPLE, FOREIGN MATTER WAS IDENTIFIED WITHIN THE TRAY. FOURIER TRANSFORM INFRARED SPECTROSCOPY (FTIR) ANALYSIS WAS PERFORMED ON THE FOREIGN MATTER. THE FTIR ANALYSIS RESULTS INDICATED THAT THE FOREIGN MATTER WAS INK. AS THE CONVENIENCE TRAY LIDS ARE PRINTED, THEY MOVE DOWN A CONVEYOR. INK RESIDUE WAS FOUND TO ACCUMULATE AT THE END OF THE PRINTING MACHINE'S CONVEYOR. INK RESIDUE WAS FOUND TO ACCUMULATE AT THE END OF THE PRINTING MACHINE'S CONVEYOR. WHEN THE LID INVOLVED IN THIS INCIDENT CAME INTO CONTACT WITH THE END OF THE CONVEYOR, IT BECAME CONTAMINATED WITH INK RESIDUE. THIS DEFECT WAS UNDETECTED DURING THE PRODUCTION PROCESS OF LOT NUMBER: 9213387, AS ALL INSPECTED PRODUCT WAS FOUND TO BE WITHIN SPECIFICATION. A PREVENTIVE MAINTENANCE PLAN HAS BEEN INITIATED TO PREVENT THIS DEFECT FROM RECURRING.

Description of Event or Problem · 0

IT WAS REPORTED THAT PRIOR TO USE FOREIGN MATTER WAS DISCOVERED IN STERILE SYRINGE TRAY WITH A BD 1 ML SYRINGE LUER SLIP TIP CONVENIENCE PAK. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: IT WAS REPORTED THAT A FOREIGN SUBSTANCE WAS FOUND IN ONE OF THE CUSTOMERS STERILE SYRINGES TRAYS.

Additional Manufacturer Narrative · 1

A SAMPLE IS AVAILABLE FOR EVALUATION. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT PRIOR TO USE FOREIGN MATTER WAS DISCOVERED IN STERILE SYRINGE TRAY WITH A BD 1 ML SYRINGE LUER SLIP TIP CONVENIENCE PAK. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: IT WAS REPORTED THAT A FOREIGN SUBSTANCE WAS FOUND IN ONE OF THE CUSTOMERS STERILE SYRINGES TRAYS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1160933 BD 1 ML SYRINGE LUER SLIP TIP CONVENIENCE PAK PISTON SYRINGE FMF BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V. 309701 9213387 30382903097013

Patients

Seq Age Sex Outcome Treatment
1 Other